Flame AAS: Zinc in Water at 213.9 nm
To determine zinc in drinking water by flame AAS at 213.9 nm.
Theoretical Foundations
Zinc's resonance line at 213.9 nm gives sensitive AAS detection down to ~0.01 mg L⁻¹. Samples are acidified to keep Zn dissolved; deuterium background correction compensates for matrix absorption at this low-UV wavelength.
Required Apparatus & Chemicals
Laboratory Apparatus
- Atomic absorption spectrometer
- Zn hollow-cathode lamp
- Deuterium background corrector
- Volumetric flasks
- Pipettes
Chemical Reagents
- Zn standard (10 mg L⁻¹)
- Nitric acid (for acidification)
- Water samples
Step-by-Step Practical Procedure
- Acidify samples to pH < 2 with HNO3.
- Prepare Zn standards 0-1 mg L⁻¹ in dilute HNO3.
- Install Zn lamp; select 213.9 nm; enable background correction.
- Optimize flame; zero with acid blank.
- Read standards and samples.
- Determine Zn and compare with the 3 mg L⁻¹ guideline.
Observations & Expected Results
Linear calibration; typical slope ≈ 0.48 A·L mg⁻¹; background correction removes matrix offsets.
Worked Stoichiometric Calculations
Problem: Slope 0.48 A·L mg⁻¹ at 213.9 nm. A sample gives A = 0.240. Find Zn in mg L⁻¹.
Working: c = 0.240/0.48 = 0.50 mg L⁻¹ Zn (below the 3 mg L⁻¹ guideline).
Safety Precautions & Hazard Control
- Acetylene safety as in all flame AAS.
- Background correction is essential at 213.9 nm.
- Acid-wash bottles to prevent Zn adsorption.
- Avoid Zn contamination from rubber and dust.
Waste Disposal Protocol
Acidified Zn waste to the aqueous heavy-metal waste bottle.
Viva Voce Oral Defense Questions & Answers
Q: Why is background correction important at 213.9 nm?
A: Many matrices absorb or scatter in the far UV; deuterium correction subtracts this non-atomic background.
Q: Why acidify the sample?
A: To keep zinc dissolved and prevent adsorption on container walls.
Q: Why is the Zn line so sensitive?
A: 213.9 nm is the ground-state resonance transition with high oscillator strength, efficiently populated in the flame.