Standard Addition in Flame Photometry: Sodium in Brine

To determine sodium in concentrated brine by flame photometric standard addition.

Level: UniversitySafety: Needs supervisionTime: 90 min
Aim of Experiment: To determine sodium in concentrated brine by flame photometric standard addition.

Theoretical Foundations

High salt matrices change nebulization efficiency and flame temperature, so external Na standards give biased slopes. Adding known Na spikes to equal brine aliquots and extrapolating emission to zero corrects the matrix effect, exactly as in spectrophotometric standard addition.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Flame photometer
  • Volumetric flasks 100 mL
  • Pipettes
  • Beakers
  • Wash bottle

Chemical Reagents

  • NaCl standard (1000 mg L⁻¹ Na)
  • Brine sample
  • Distilled water

Step-by-Step Practical Procedure

  1. Dilute brine 1:50 (still a strong matrix).
  2. Add 10 mL diluted brine to four flasks; spike 0, 5, 10, 15 mL of 100 mg L⁻¹ Na standard; dilute to 100 mL.
  3. Set 589 nm filter; stabilize flame.
  4. Read emissions.
  5. Plot emission versus added Na.
  6. Extrapolate to zero emission for the brine Na.

Observations & Expected Results

Emission rises linearly, e.g. 30, 45, 60, 75 divisions; negative x-intercept gives the diluted-brine Na.

Worked Stoichiometric Calculations

Problem: Added Na: 0, 5, 10, 15 mg L⁻¹ give emissions 30, 45, 60, 75 divisions. The brine was diluted 1:50. Find Na in the original brine.

Working: Slope = 15/5 = 3.0 div·L mg⁻¹. x-intercept = −30/3.0 = −10 mg L⁻¹. Diluted brine = 10 mg L⁻¹; original = 10 × 50 = 500 mg L⁻¹ Na.

Safety Precautions & Hazard Control

  • Keep the matrix identical in all flasks (equal brine aliquots).
  • Spikes must not overwhelm the matrix.
  • Verify linearity over the spiked range.
  • Rinse the nebulizer between readings.

Waste Disposal Protocol

Dilute brine to the aqueous drain.

Viva Voce Oral Defense Questions & Answers

Q: Why does brine need standard addition?

A: Its high dissolved solids alter droplet formation and flame temperature, changing sensitivity versus pure-water standards.

Q: Why must every flask contain the same brine volume?

A: So the matrix effect is identical for sample and spikes and cancels in the extrapolation.

Q: What does the x-intercept represent?

A: The added sodium that would exactly cancel the sample's own sodium signal — equal in magnitude to the sample concentration.