CHEMVEN

Clinical Flame Photometry: Sodium and Potassium in Serum

To determine serum Na⁺ and K⁺ by flame photometry against clinical standards.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To determine serum Na⁺ and K⁺ by flame photometry against clinical standards.

Theoretical Foundations

Serum diluted 1:200 is aspirated; Na read at 589 nm, K at 766.5 nm. Internal standardization with Li compensates flame fluctuations. Results in mmol L⁻¹ are compared with reference ranges (Na 135-145, K 3.5-5.0 mmol L⁻¹).

Required Apparatus & Chemicals

Laboratory Apparatus

  • Flame photometer (clinical, with Li internal standard)
  • Micropipettes
  • Volumetric flasks
  • Test tubes
  • Vortex mixer

Chemical Reagents

  • Mixed Na/K standard (140/4.0 mmol L⁻¹)
  • Lithium internal standard diluent
  • Serum (control serum)

Step-by-Step Practical Procedure

  1. Dilute serum 1:200 in Li diluent.
  2. Dilute the mixed standard identically.
  3. Set Na filter; read standard then samples.
  4. Set K filter; repeat.
  5. Calculate using the standard ratio.
  6. Compare with reference ranges.

Observations & Expected Results

Stable readings with Li internal standard; serum Na ≈ 140 mmol L⁻¹, K ≈ 4.0 mmol L⁻¹ in controls.

Worked Stoichiometric Calculations

Problem: The 140 mmol L⁻¹ Na standard reads 70 divisions; a serum reads 70 divisions. The 4.0 mmol L⁻¹ K standard reads 35 divisions; the serum reads 35 divisions. Report serum Na and K.

Working: Na = 70/70 × 140 = 140 mmol L⁻¹. K = 35/35 × 4.0 = 4.0 mmol L⁻¹. Both within reference ranges.

Safety Precautions & Hazard Control

  • Handle serum with gloves (biohazard).
  • Use the Li internal standard to cancel flame drift.
  • Dilute accurately; the 1:200 factor multiplies errors.
  • Deproteinize if the burner clogs.

Waste Disposal Protocol

Serum waste to biohazard disposal; dilute salts to drain.

Viva Voce Oral Defense Questions & Answers

Q: What is the purpose of the lithium internal standard?

A: Li emission tracks flame fluctuations; ratioing Na/Li and K/Li cancels drift and nebulizer variation.

Q: Why dilute serum 1:200?

A: To reach the linear flame range and to minimize viscosity and protein matrix effects.

Q: Why measure Na and K together clinically?

A: Their balance reflects fluid/electrolyte status; abnormal ratios indicate renal or adrenal disorders.

← Back to Experiments Directory