Flame Photometry: Sodium in Water at 589 nm
To determine sodium in drinking water by flame photometry at 589 nm.
Theoretical Foundations
In a flame, Na atoms are thermally excited and emit the characteristic yellow doublet at 589 nm. Emission intensity is proportional to sodium concentration over 0-10 mg L⁻¹. A calibration of emission versus standards converts the sample reading to mg L⁻¹ Na.
Required Apparatus & Chemicals
Laboratory Apparatus
- Flame photometer
- Volumetric flasks 100 mL
- Pipettes
- Beakers
- Gas supply (LPG/air)
Chemical Reagents
- NaCl standard (100 mg L⁻¹ Na)
- Distilled water
- Unknown water sample
Step-by-Step Practical Procedure
- Prepare Na standards 0-10 mg L⁻¹.
- Light the flame photometer; allow 15 min stabilization.
- Set the 589 nm filter; zero with distilled water.
- Aspirate standards; record emission readings.
- Aspirate the unknown sample.
- Read Na from the calibration curve.
Observations & Expected Results
Yellow flame colour intensifies with Na; emission linear with concentration; typical slope ≈ 8.5 scale divisions·L mg⁻¹.
Worked Stoichiometric Calculations
Problem: Slope 8.5 divisions·L mg⁻¹. A sample reads 42.5 divisions. Find Na in mg L⁻¹.
Working: c = 42.5/8.5 = 5.00 mg L⁻¹ Na.
Safety Precautions & Hazard Control
- Stabilize the flame before reading.
- Rinse the nebulizer with water between samples.
- Keep gas pressures constant.
- Sodium contamination is ubiquitous; run blanks.
Waste Disposal Protocol
Dilute salt solutions to the aqueous drain.
Viva Voce Oral Defense Questions & Answers
Q: Why is flame photometry emission, not absorption?
A: The flame thermally excites atoms; we measure the light they emit on returning to ground state, at 589 nm for Na.
Q: Why zero with distilled water?
A: To subtract flame background emission and any sodium contamination in the water.
Q: What causes the yellow colour?
A: The Na 3p → 3s electronic transition (D lines) at 589.0/589.6 nm.