Precision Study: Repeatability of Absorbance Readings

To evaluate instrumental precision from replicate absorbance measurements and express it as RSD.

Level: UniversitySafety: SafeTime: 45 min
Aim of Experiment: To evaluate instrumental precision from replicate absorbance measurements and express it as RSD.

Theoretical Foundations

Precision is quantified by the relative standard deviation (RSD = s/mean × 100%) of n replicate readings of one stable solution. Good UV-Vis instruments give RSD < 1% at mid-range absorbance. This experiment separates random instrument noise from sample preparation error.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV-Vis spectrophotometer
  • Volumetric flask
  • Pipette
  • Cuvette

Chemical Reagents

  • KMnO4 solution (~0.5 absorbance units)
  • KMnO4 (AR grade, for stock)
  • Distilled water (diluent)

Step-by-Step Practical Procedure

  1. Prepare one KMnO4 solution (A ≈ 0.500 at 525 nm).
  2. Zero with water; read the same solution 10 times, removing and replacing the cuvette each time.
  3. Record all 10 absorbances.
  4. Calculate mean and standard deviation.
  5. Compute RSD.
  6. Compare with the <1% criterion.

Observations & Expected Results

Readings cluster tightly, e.g. 0.498-0.503; RSD typically 0.3-0.8%.

Worked Stoichiometric Calculations

Problem: Ten readings give mean 0.500 and SD 0.003. Find the RSD and judge precision.

Working: RSD = 0.003/0.500 × 100 = 0.60%. Below 1%: good precision.

Safety Precautions & Hazard Control

  • Use one solution to isolate instrument precision.
  • Re-seat the cuvette each time to include positioning error.
  • Allow lamp warm-up before starting.
  • Avoid touching cuvette optical faces.

Waste Disposal Protocol

Dilute permanganate to oxidizer waste.

Viva Voce Oral Defense Questions & Answers

Q: Precision versus accuracy: distinguish them.

A: Precision is reproducibility (tight clustering); accuracy is closeness to the true value. Good precision does not guarantee accuracy.

Q: Why re-seat the cuvette?

A: To include real-world positioning variability, not just electronic noise.

Q: What dominates noise at very low absorbance?

A: Detector and lamp noise relative to the tiny absorbed signal; hence the 0.1-1.0 working range.