CHEMVEN

Photometric Accuracy: Potassium Dichromate Absorbance Standards

To check photometric (absorbance) accuracy using certified K2Cr2O7 solutions.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To check photometric (absorbance) accuracy using certified K2Cr2O7 solutions.

Theoretical Foundations

Acidic K2Cr2O7 solutions have internationally tabulated absorptivities (e.g. at 350 nm). Measuring a prepared standard and comparing with the certified absorbance tests the photometric scale; errors should be within ±1% for quantitative work.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV-Vis spectrophotometer
  • Volumetric flasks
  • Analytical balance
  • Quartz cuvettes

Chemical Reagents

  • Potassium dichromate (analytical grade, dried)
  • Sulfuric acid (0.01 M)
  • Distilled water (diluent)

Step-by-Step Practical Procedure

  1. Dry K2Cr2O7 at 120 °C; weigh accurately.
  2. Prepare 60.0 mg L⁻¹ in 0.01 M H2SO4.
  3. Zero at 350 nm with 0.01 M H2SO4.
  4. Read the absorbance in triplicate.
  5. Compare with the certified value (0.640).
  6. Calculate % photometric error.

Observations & Expected Results

Stable yellow solution; certified A350 = 0.640 for 60.0 mg L⁻¹; well-maintained instruments read within ±0.006.

Worked Stoichiometric Calculations

Problem: Certified A350 = 0.640. The instrument reads 0.635. Find the % photometric error and judge acceptability (±1%).

Working: Error = (0.635 − 0.640)/0.640 × 100 = −0.78%. Within ±1%: acceptable.

Safety Precautions & Hazard Control

  • K2Cr2O7 is carcinogenic; weigh carefully with minimal dust.
  • Dry the salt; moisture biases the mass.
  • Use the specified 0.01 M H2SO4 medium.
  • Read promptly; dichromate is stable but dust settles.

Waste Disposal Protocol

Dichromate standard to the chromium waste bottle.

Viva Voce Oral Defense Questions & Answers

Q: Why is K2Cr2O7 a photometric standard?

A: It is stable, pure, weighable, and its absorptivity is internationally certified at several wavelengths.

Q: Why the specified sulfuric acid medium?

A: The dichromate/chromate equilibrium (and hence the spectrum) depends on pH; the certified values assume 0.01 M H2SO4.

Q: What causes photometric error?

A: Stray light, dirty cuvettes, lamp drift and detector nonlinearity.