Solvent Cutoff Study: Usable UV Range of Common Solvents

To determine the UV cutoff wavelengths of common solvents and choose correct solvents for UV work.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To determine the UV cutoff wavelengths of common solvents and choose correct solvents for UV work.

Theoretical Foundations

Every solvent absorbs below a cutoff wavelength (A = 1.0 for 1 cm): water 190, ethanol 210, hexane 210, chloroform 245, acetone 330 nm. Measurements below the cutoff are meaningless because the solvent, not the sample, absorbs the beam.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV spectrophotometer
  • Quartz cuvettes
  • Solvent bottles
  • Pasteur pipettes
  • Wash bottle

Chemical Reagents

  • Water
  • Ethanol
  • Hexane
  • Chloroform
  • Acetone (HPLC grade)

Step-by-Step Practical Procedure

  1. Fill a quartz cuvette with each solvent; blank with air.
  2. Scan 200-400 nm for each solvent.
  3. Note the wavelength where A reaches 1.0 (cutoff).
  4. Tabulate cutoffs and compare with literature.
  5. Decide which solvents suit a 220 nm analysis.
  6. Repeat for the solvent you plan to use in later experiments.

Observations & Expected Results

Water and hexane transmit to ~200 nm; acetone is opaque below ~330 nm; each solvent shows a sharp absorption edge.

Worked Stoichiometric Calculations

Problem: An acetone blank gives A = 1.5 at 200 nm in a 1 cm cell. Can acetone be used as solvent for a 220 nm determination?

Working: No. Acetone's cutoff is ~330 nm; at 220 nm the solvent absorbs essentially all light, leaving no measurable sample signal.

Safety Precautions & Hazard Control

  • Use quartz cuvettes for UV scans.
  • Keep cuvettes scrupulously clean; residues absorb in the UV.
  • Work in a fume cupboard with volatile solvents.
  • Blank with air, not with another solvent.

Waste Disposal Protocol

Return solvents to their bottles or to organic waste as appropriate.

Viva Voce Oral Defense Questions & Answers

Q: What is a solvent cutoff?

A: The wavelength below which the pure solvent itself absorbs too strongly (A > 1) for analytical measurements.

Q: Why is acetone a poor UV solvent?

A: Its n→π* carbonyl absorption sets the cutoff near 330 nm, blocking the whole mid-UV region.

Q: Which solvents suit far-UV work near 200 nm?

A: Water, hexane and cyclohexane, with cutoffs near 190-210 nm.