First-Order Kinetics: Catalyzed Decomposition of H2O2 at 240 nm

To verify first-order kinetics for iodide-catalyzed H2O2 decomposition by UV at 240 nm.

Level: UniversitySafety: Needs supervisionTime: 90 min
Aim of Experiment: To verify first-order kinetics for iodide-catalyzed H2O2 decomposition by UV at 240 nm.

Theoretical Foundations

H2O2 absorbs at 240 nm while water does not, so A240 tracks [H2O2] directly. For first-order decay, ln(A_t) versus t is linear with slope −k. The half-life is concentration-independent, confirming first order.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV spectrophotometer
  • Quartz cuvettes
  • Pipettes
  • Stopwatch

Chemical Reagents

  • Hydrogen peroxide (0.10 M)
  • Potassium iodide catalyst (0.01 M)
  • Distilled water

Step-by-Step Practical Procedure

  1. Mix H2O2 with a trace of KI catalyst; start timing.
  2. Record A240 every 2 min for 30 min.
  3. Plot ln(A) versus time.
  4. Determine k from the slope.
  5. Calculate t½ and check it is constant between successive half-lives.
  6. Repeat without catalyst to show catalysis.

Observations & Expected Results

A240 decays exponentially; ln plot linear; typical k ≈ 0.035 min⁻¹ with catalyst.

Worked Stoichiometric Calculations

Problem: A240 falls from 0.800 to 0.400 in 20.0 min. Assuming first order, find k.

Working: One half-life = 20.0 min (0.800 → 0.400). k = 0.693/t½ = 0.693/20.0 = 0.0347 min⁻¹.

Safety Precautions & Hazard Control

  • H2O2 is corrosive; wear gloves and goggles.
  • Use quartz cuvettes for 240 nm.
  • Keep catalyst concentration identical between runs.
  • Avoid transition-metal contamination which catalyzes erratically.

Waste Disposal Protocol

Dilute peroxide/iodide waste to the aqueous drain.

Viva Voce Oral Defense Questions & Answers

Q: How does constant half-life prove first order?

A: Only first-order reactions have t½ = 0.693/k independent of initial concentration.

Q: What is the catalyst's role?

A: I⁻ provides a lower-energy pathway (via HIO intermediate) without being consumed.

Q: Why quartz cuvettes?

A: Glass absorbs at 240 nm; quartz is transparent in the UV.