Simultaneous Permanganate and Dichromate: Two-Wavelength Analysis
To resolve a KMnO4/K2Cr2O7 mixture by simultaneous equations at 525 and 440 nm.
Theoretical Foundations
MnO4⁻ absorbs strongly at 525 nm and weakly at 440 nm, while Cr2O7²⁻ absorbs strongly at 440 nm and weakly at 525 nm. With the four molar absorptivities known, the two measured absorbances give two linear equations for the two oxidant concentrations, a classic analytical application of matrix methods.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks
- Pipettes
- Cuvettes
Chemical Reagents
- KMnO4 standard
- K2Cr2O7 standard
- Unknown mixture
- Dilute H2SO4
Step-by-Step Practical Procedure
- Determine ε of MnO4⁻ and Cr2O7²⁻ at 525 and 440 nm from pure standards.
- Read A525 and A440 of the unknown mixture.
- Write the two additivity equations.
- Solve the 2 × 2 system for both concentrations.
- Check the result by mass balance or a third wavelength.
- Report each oxidant in mol L⁻¹.
Observations & Expected Results
Mixture is orange-purple; A525 dominated by permanganate, A440 by dichromate; the solution is well-conditioned for simultaneous analysis.
Worked Stoichiometric Calculations
Problem: ε: MnO4⁻: 2400 (525), 400 (440); Cr2O7²⁻: 300 (525), 1800 (440). Mixture: A525 = 0.720, A440 = 0.560. Find both concentrations.
Working: 2400m + 300c = 0.720; 400m + 1800c = 0.560. Multiply second by 6: 2400m + 10800c = 3.360. Subtract: 10500c = 2.640 → c(Cr2O7²⁻) = 2.51 × 10⁻⁴ mol L⁻¹. Then m = (0.720 − 300 × 2.51 × 10⁻⁴)/2400 = 2.69 × 10⁻⁴ mol L⁻¹ MnO4⁻.
Safety Precautions & Hazard Control
- Keep both oxidants in the same acidic medium.
- Permanganate slowly oxidises organics; read promptly.
- Dichromate is carcinogenic; wear gloves.
- Verify linearity for each component separately first.
Waste Disposal Protocol
Mixed Cr/Mn oxidant waste to the chromium waste bottle after reduction.
Viva Voce Oral Defense Questions & Answers
Q: Why is this pair ideal for two-wavelength analysis?
A: Each ion dominates a different wavelength, so the equations are nearly independent and errors stay small.
Q: What matrix condition guarantees a unique solution?
A: A non-zero determinant of the absorptivity matrix, i.e. the two spectra must not be proportional.
Q: How would a third wavelength help?
A: It provides an overdetermined check: the third predicted absorbance must match the measured one, validating the analysis.