Two-Component Analysis: Simultaneous Cr and Co Determination
To determine Cr(III) and Co(II) in a mixture without separation using absorbances at two wavelengths.
Theoretical Foundations
When spectra overlap, absorbances are additive: A(λ1) = εCr1·cCr + εCo1·cCo and A(λ2) = εCr2·cCr + εCo2·cCo. Measuring at two wavelengths where the absorptivity ratios differ gives two equations solved simultaneously for both concentrations, provided Beer's law holds for each component.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks
- Pipettes
- Cuvettes
Chemical Reagents
- Cr(III) standard
- Co(II) standard
- Unknown Cr/Co mixture
Step-by-Step Practical Procedure
- Record spectra of pure Cr(III) and Co(II); select λ1 = 410 nm (Cr max) and λ2 = 510 nm (Co max).
- Determine the four molar absorptivities from pure standards.
- Read A410 and A510 of the unknown mixture.
- Set up the two simultaneous equations.
- Solve for cCr and cCo.
- Verify by checking a third wavelength.
Observations & Expected Results
Cr(III) is violet-blue, Co(II) pink; the mixture spectrum is the sum; chosen wavelengths maximize the determinant of the coefficient matrix.
Worked Stoichiometric Calculations
Problem: At 410 nm: A = 0.680 = 3200cCr + 120cCo. At 510 nm: A = 0.540 = 60cCr + 480cCo (ε in L mol⁻¹ cm⁻¹, l = 1 cm). Solve for both.
Working: From the first: cCr = (0.680 − 120cCo)/3200. Substituting: 60(0.680 − 120cCo)/3200 + 480cCo = 0.540 → 0.01275 − 2.25cCo + 480cCo = 0.540 → cCo = 1.10 × 10⁻³ mol L⁻¹. Then cCr = (0.680 − 0.132)/3200 = 1.71 × 10⁻⁴ mol L⁻¹.
Safety Precautions & Hazard Control
- Choose wavelengths where absorptivity ratios differ greatly.
- Ensure no chemical interaction between the components.
- Work in the linear absorbance range for both species.
- Use matched cuvettes for all readings.
Waste Disposal Protocol
Dilute Cr/Co solutions to the aqueous heavy-metal waste bottle.
Viva Voce Oral Defense Questions & Answers
Q: What is the key assumption of simultaneous analysis?
A: Additivity: total absorbance equals the sum of individual absorbances, i.e. the components do not interact chemically.
Q: Why choose wavelengths of maximum difference?
A: It maximizes the determinant of the equations, minimizing the propagation of measurement error into the concentrations.
Q: When does this method fail?
A: If spectra are too similar (determinant near zero) or if the components react with each other.