CHEMVEN

Creatinine Determination: Jaffe Reaction at 520 nm

To determine creatinine in serum by the alkaline picrate (Jaffe) reaction at 520 nm.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To determine creatinine in serum by the alkaline picrate (Jaffe) reaction at 520 nm.

Theoretical Foundations

Creatinine reacts with picric acid in alkali to form an orange-red Janovsky complex (λmax 520 nm). The reaction is not fully specific (glucose, acetone react slowly), so absorbance is read at a fixed time (kinetic Jaffe) or after deproteinization. The clinical range is ~0.6-1.2 mg dL⁻¹.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV-Vis spectrophotometer
  • Test tubes
  • Centrifuge
  • Pipettes
  • Stopwatch
  • Cuvettes

Chemical Reagents

  • Creatinine standard (2 mg dL⁻¹)
  • Picric acid solution
  • Sodium hydroxide (0.75 M)
  • Trichloroacetic acid (10%)
  • Serum sample

Step-by-Step Practical Procedure

  1. Deproteinize serum with TCA; centrifuge.
  2. Prepare creatinine standards 0-2 mg dL⁻¹.
  3. Mix 1 mL sample/standard with 1 mL picric acid and 1 mL NaOH.
  4. Start timer; read A520 exactly 10 min after mixing.
  5. Zero with the reagent blank prepared identically.
  6. Determine creatinine from the calibration curve.

Observations & Expected Results

Orange-red colour develops; A520 linear with creatinine; typical slope ≈ 0.25 A·dL mg⁻¹.

Worked Stoichiometric Calculations

Problem: Slope 0.25 A·dL mg⁻¹ at 520 nm. A serum filtrate gives A = 0.300. Find creatinine in mg dL⁻¹.

Working: c = 0.300/0.25 = 1.20 mg dL⁻¹ (upper end of the normal range).

Safety Precautions & Hazard Control

  • Read at exactly the fixed time; slow interferents add colour later.
  • Picric acid must stay moist; dry picric acid is explosive.
  • Deproteinize to remove slow-reacting interferents.
  • Keep NaOH concentration identical in all tubes.

Waste Disposal Protocol

Keep picric acid waste moist and collect in the designated picrate waste bottle.

Viva Voce Oral Defense Questions & Answers

Q: Why read at a fixed time?

A: Creatinine reacts fast while interferents (glucose, ketones) react slowly; a fixed-time reading minimizes their contribution.

Q: Why is dry picric acid dangerous?

A: Dry picric acid and its salts are shock-sensitive explosives; it must always be kept wetted.

Q: What does elevated creatinine indicate clinically?

A: Reduced kidney filtration (glomerular filtration rate), since creatinine is cleared by the kidneys.

← Back to Experiments Directory