Nickel in Alloy Steel: DMG Extraction Method
To determine % nickel in alloy steel by DMG extraction into chloroform at 465 nm.
Theoretical Foundations
After dissolving steel in acid, nickel is selectively extracted as red Ni(DMG)2 into chloroform (465 nm, ε ≈ 1.7 × 10⁴), leaving iron and chromium in the aqueous phase. The extraction both separates and concentrates nickel, allowing determination of tenths of a percent Ni in steel.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Separating funnels
- Hot plate
- Volumetric flasks
- Pipettes
- Cuvettes
Chemical Reagents
- Alloy steel (1.000 g)
- HCl/HNO3 for dissolution
- Dimethylglyoxime (1% in ethanol)
- Ammonia (1:1)
- Bromine water
- Chloroform
- Ni standard
Step-by-Step Practical Procedure
- Dissolve 1.000 g steel in HCl/HNO3; dilute to 100 mL.
- Take the full 100 mL (or a measured aliquot) to a separating funnel.
- Add bromine water, then ammonia until ammoniacal.
- Add DMG; extract twice with 10 mL chloroform.
- Combine extracts; dilute to 50 mL with chloroform.
- Extract Ni standards identically; zero at 465 nm with CHCl3; read and calculate % Ni.
Observations & Expected Results
Red complex concentrates in the chloroform layer; aqueous iron stays behind; A465 linear with nickel.
Worked Stoichiometric Calculations
Problem: 1.000 g steel to 100 mL; the whole digest extracted to 50 mL CHCl3 gives A = 0.350 at 465 nm (ε = 1.70 × 10⁴, l = 1 cm). Find % Ni.
Working: c = 0.350/1.70 × 10⁴ = 2.06 × 10⁻⁵ mol L⁻¹. Mass in 0.050 L = 2.06 × 10⁻⁵ × 0.050 × 58.69 = 6.04 × 10⁻⁵ g. %Ni = 6.04 × 10⁻⁵/1.000 × 100 = 0.121%.
Safety Precautions & Hazard Control
- Extract standards through the identical procedure to cancel extraction losses.
- Vent the funnel when shaking chloroform.
- Work in a fume cupboard with chloroform.
- Keep the aqueous phase ammoniacal during extraction.
Waste Disposal Protocol
Chloroform layers to halogenated waste; acidic aqueous iron waste to heavy-metal waste.
Viva Voce Oral Defense Questions & Answers
Q: Why extract the standards too?
A: Extraction is never 100% efficient; identical treatment of standards cancels the recovery factor.
Q: Why does iron not extract with DMG?
A: Fe does not form a neutral chloroform-soluble DMG chelate under these conditions; it stays in the ammoniacal aqueous phase.
Q: What is the advantage of extraction over direct colorimetry?
A: It separates nickel from the intensely coloured iron matrix and concentrates it into a small organic volume, boosting sensitivity.