Nitrite in Water: Griess Method for Environmental Monitoring
To monitor nitrite in surface water by the Griess method at 540 nm against the 1 mg L⁻¹ guideline.
Theoretical Foundations
The Griess diazotization-coupling sequence gives the pink azo dye (540 nm) specific for nitrite, with detection down to ~0.01 mg L⁻¹. Nitrite is an intermediate in nitrification and indicates recent organic pollution or fertilizer runoff, so rapid analysis after sampling matters.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks 50 mL
- Pipettes
- Cuvettes
- Sample bottles
Chemical Reagents
- NaNO2 standard (1 mg L⁻¹)
- Sulfanilamide reagent
- NED reagent
- River/well water sample
Step-by-Step Practical Procedure
- Collect water in clean bottles; analyse within 6 h.
- Prepare nitrite standards 0-0.5 mg L⁻¹ in 50 mL flasks.
- Add sulfanilamide, wait 5 min, add NED to standards and sample.
- Wait 15 min; zero at 540 nm with blank.
- Read absorbances; plot the calibration curve.
- Determine nitrite and compare with the 1 mg L⁻¹ guideline.
Observations & Expected Results
Pink dye develops; natural waters usually read below 0.5 mg L⁻¹; polluted drains read higher.
Worked Stoichiometric Calculations
Problem: Slope 1.05 A·L mg⁻¹; a river sample gives A = 0.262 at 540 nm. Find nitrite in mg L⁻¹.
Working: c = 0.262/1.05 = 0.250 mg L⁻¹ NO2⁻, below the 1 mg L⁻¹ guideline.
Safety Precautions & Hazard Control
- Analyse quickly; nitrite oxidises to nitrate in stored samples.
- Refrigerate samples if analysis is delayed.
- Remove turbidity by filtration.
- Nitrite standards are unstable; prepare fresh.
Waste Disposal Protocol
Azo-dye waste to the organic-aqueous waste container.
Viva Voce Oral Defense Questions & Answers
Q: Why does nitrite indicate recent pollution?
A: Nitrite is a short-lived intermediate between ammonia and nitrate; its presence means nitrification is actively occurring, often from fresh sewage or fertilizer.
Q: Why analyse within hours of sampling?
A: Dissolved oxygen and bacteria convert nitrite to nitrate, so delayed analysis underestimates the original level.
Q: How would you determine nitrate plus nitrite together?
A: Reduce nitrate to nitrite on a cadmium column, run the Griess method for the sum, and subtract the separately measured nitrite.