Chromium(VI) in Wastewater: Diphenylcarbazide Trace Analysis
To quantify Cr(VI) in industrial effluent by diphenylcarbazide at 540 nm and compare with discharge limits.
Theoretical Foundations
The diphenylcarbazide reaction gives the red-violet Cr complex (540 nm, ε ≈ 4.2 × 10⁴), detecting Cr(VI) down to ~0.01 mg L⁻¹. Effluents are filtered and pH-adjusted; only Cr(VI) responds, so a separate oxidation step would be needed for total chromium. Results are compared with the typical 0.1 mg L⁻¹ discharge limit.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Filtration assembly
- Volumetric flasks 100 mL
- Pipettes
- Cuvettes
- pH meter
Chemical Reagents
- K2Cr2O7 standard (1 mg L⁻¹ as Cr)
- Diphenylcarbazide reagent
- Sulfuric acid (1 M)
- NaOH for pH adjustment
- Effluent sample
Step-by-Step Practical Procedure
- Filter the effluent through 0.45 µm membrane.
- Adjust an aliquot to pH ~2 with H2SO4.
- Prepare Cr(VI) standards 0-0.5 mg L⁻¹.
- Add diphenylcarbazide to standards and sample; wait 10 min.
- Zero at 540 nm with blank; read absorbances.
- Calculate Cr(VI) and compare with the 0.1 mg L⁻¹ limit.
Observations & Expected Results
Red-violet colour proportional to Cr(VI); effluent samples often read near or above the discharge limit without treatment.
Worked Stoichiometric Calculations
Problem: An effluent gives A = 0.340 at 540 nm (ε = 4.20 × 10⁴, l = 1 cm). Find Cr(VI) in mg L⁻¹ and judge against a 0.1 mg L⁻¹ limit.
Working: c = 0.340/4.20 × 10⁴ = 8.10 × 10⁻⁶ mol L⁻¹ = 8.10 × 10⁻⁶ × 52.00 × 1000 = 0.421 mg L⁻¹. This exceeds the 0.1 mg L⁻¹ discharge limit.
Safety Precautions & Hazard Control
- Cr(VI) is carcinogenic; wear gloves throughout.
- Filter turbid effluents; particles scatter at 540 nm.
- Analyse Cr(VI) promptly; it reduces to Cr(III) in organic-rich effluents.
- Do not acidify excessively; the reagent needs mild acid.
Waste Disposal Protocol
All chromium waste to the dedicated chromium waste bottle.
Viva Voce Oral Defense Questions & Answers
Q: Does this method give total chromium?
A: No, only Cr(VI). Total chromium needs prior oxidation of Cr(III), e.g. with persulfate.
Q: Why filter at 0.45 µm?
A: Suspended solids scatter light at 540 nm and would falsely elevate the reading; filtration isolates dissolved Cr(VI).
Q: Why is Cr(VI) regulated more strictly than Cr(III)?
A: Cr(VI) is carcinogenic and highly mobile in water, while Cr(III) is an essential trace nutrient of low toxicity.