Iron in Drinking Water: Phenanthroline Trace Method

To determine trace iron in drinking water against the 0.3 mg L⁻¹ guideline using 1,10-phenanthroline.

Level: UniversitySafety: Needs supervisionTime: 90 min
Aim of Experiment: To determine trace iron in drinking water against the 0.3 mg L⁻¹ guideline using 1,10-phenanthroline.

Theoretical Foundations

The phenanthroline method (λmax 510 nm, ε = 1.11 × 10⁴) detects iron down to ~0.02 mg L⁻¹, well below the aesthetic guideline of 0.3 mg L⁻¹. Hydroxylamine reduces all iron to Fe(II) and acetate buffer holds pH 4.5, so the measured [Fe(phen)3]²⁺ represents total dissolved iron.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV-Vis spectrophotometer
  • Volumetric flasks 100 mL
  • Pipettes
  • Cuvettes
  • Sample bottles (acid-washed)

Chemical Reagents

  • Fe standard (1 mg L⁻¹)
  • Hydroxylamine hydrochloride
  • Acetate buffer (pH 4.5)
  • 1,10-phenanthroline
  • Tap/well water sample

Step-by-Step Practical Procedure

  1. Acidify the water sample at collection to keep iron dissolved.
  2. Prepare standards 0-0.5 mg L⁻¹ Fe in 100 mL flasks.
  3. Add hydroxylamine, buffer and phenanthroline to standards and sample.
  4. Wait 15 min; zero at 510 nm with the blank.
  5. Read absorbances; build the calibration curve.
  6. Read the sample iron and compare with the 0.3 mg L⁻¹ guideline.

Observations & Expected Results

Faint orange colour at trace levels; calibration linear; typical drinking water gives A510 = 0.05-0.30.

Worked Stoichiometric Calculations

Problem: A water sample gives A = 0.280 at 510 nm (ε = 1.11 × 10⁴, l = 1 cm, no dilution). Find Fe in mg L⁻¹ and state if it exceeds 0.3 mg L⁻¹.

Working: c = 0.280/1.11 × 10⁴ = 2.52 × 10⁻⁵ mol L⁻¹ = 2.52 × 10⁻⁵ × 55.85 × 1000 = 1.41 mg L⁻¹. This exceeds the 0.3 mg L⁻¹ guideline.

Safety Precautions & Hazard Control

  • Acid-wash sample bottles; iron adsorbs on glass.
  • Analyse promptly or acidify; Fe(II) oxidises and precipitates.
  • Run a field blank to detect contamination.
  • Match the matrix of standards and samples.

Waste Disposal Protocol

Collect phenanthroline waste in the organic-aqueous waste bottle.

Viva Voce Oral Defense Questions & Answers

Q: Why acidify the sample at collection?

A: To keep iron dissolved and prevent adsorption on bottle walls and oxidation-precipitation of Fe(III).

Q: What does exceeding 0.3 mg L⁻¹ imply?

A: It is an aesthetic (taste/staining) guideline, not a health limit; high iron stains laundry and gives metallic taste.

Q: Why is the detection limit important here?

A: The method must quantify reliably below the guideline value to judge compliance; ε = 1.11 × 10⁴ provides that sensitivity.