Caffeine in Tea: UV Extraction Method at 273 nm
To determine caffeine in tea leaves by chloroform extraction and UV measurement at 273 nm.
Theoretical Foundations
Caffeine is extracted from brewed tea into chloroform after basification (which keeps tannins ionized in water). Caffeine absorbs at 273 nm (ε ≈ 9.7 × 10³ L mol⁻¹ cm⁻¹). Absorbance of the chloroform extract at 273 nm, corrected by calibration, gives caffeine mass and hence % w/w in the leaves.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV spectrophotometer
- Separating funnel
- Quartz cuvettes
- Volumetric flasks
- Water bath
- Balance
Chemical Reagents
- Tea leaves (2.0 g)
- Caffeine standard
- Chloroform
- Sodium carbonate solution
- Distilled water
Step-by-Step Practical Procedure
- Brew 2.0 g tea in 250 mL boiling water for 10 min; filter.
- Basify the brew with Na2CO3; extract 100 mL with 3 × 20 mL chloroform.
- Combine extracts, dry over Na2SO4, dilute to 100 mL.
- Dilute 10.0 mL of extract to 100 mL with chloroform.
- Prepare caffeine standards in chloroform (0-20 mg L⁻¹).
- Zero at 273 nm with chloroform; read and calculate % caffeine.
Observations & Expected Results
Chloroform layer carries the caffeine; A273 of the diluted extract is linear with caffeine; tannins stay in the alkaline aqueous layer.
Worked Stoichiometric Calculations
Problem: 2.0 g tea brewed to 250 mL; 10 mL extract diluted to 100 mL gives A = 0.420 at 273 nm (ε = 9700, l = 1 cm). Find % w/w caffeine.
Working: c(measured) = 0.420/9700 = 4.33 × 10⁻⁵ mol L⁻¹ = 4.33 × 10⁻⁵ × 194.19 × 1000 = 8.41 mg L⁻¹. Extract conc. = 84.1 mg L⁻¹. In 0.250 L: 21.0 mg. % = 0.0210/2.0 × 100 = 1.05% w/w.
Safety Precautions & Hazard Control
- Vent the separating funnel frequently during extraction.
- Basify before extraction to keep tannins aqueous.
- Use chloroform in a fume cupboard.
- Dry the extract; water droplets scatter UV light.
Waste Disposal Protocol
Collect chloroform extracts in the halogenated waste bottle.
Viva Voce Oral Defense Questions & Answers
Q: Why basify the tea brew before extraction?
A: Tannins and acids stay ionized in alkaline water and do not extract, while neutral caffeine partitions into chloroform, giving selectivity.
Q: Why is caffeine measured at 273 nm?
A: It is the π→π* absorption maximum of the purine ring system, giving maximum sensitivity.
Q: How would residual water in the extract affect the reading?
A: Water droplets scatter light and caffeine partitions partly into them, causing noisy, low absorbance readings.