Paracetamol Assay: Direct UV Measurement at 243 nm
To assay paracetamol tablets by direct UV absorbance at 243 nm using the specific absorbance A(1%, 1 cm).
Theoretical Foundations
Paracetamol shows a strong π→π* band at 243 nm in dilute acid with A(1%, 1 cm) = 715, meaning a 1 g/100 mL solution gives A = 715. No colour-forming reagent is needed; concentration follows directly from c(g/100 mL) = A/715 after suitable dilution, provided excipients do not absorb at 243 nm.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV spectrophotometer
- Quartz cuvettes
- Volumetric flasks 1000, 100 mL
- Pipettes
- Analytical balance
Chemical Reagents
- Paracetamol reference standard
- Hydrochloric acid (0.1 M)
- Paracetamol tablets (500 mg label)
Step-by-Step Practical Procedure
- Weigh powder equivalent to one 500 mg tablet; dissolve in 0.1 M HCl and dilute to 1000 mL.
- Filter; dilute 1.00 mL of filtrate to 100 mL.
- Prepare a standard check solution the same way.
- Zero at 243 nm with 0.1 M HCl using quartz cuvettes.
- Read absorbances of standard and sample.
- Calculate mg per tablet using A(1%, 1 cm) = 715.
Observations & Expected Results
Strong UV band at 243 nm; sample absorbance typically 0.3-0.4 after the 1:100 dilution.
Worked Stoichiometric Calculations
Problem: A 500 mg tablet dissolved to 1000 mL, then 1.00 mL diluted to 100 mL, gives A = 0.358 at 243 nm. Using A(1%,1cm) = 715, find mg paracetamol per tablet.
Working: c(measured) = 0.358/715 = 5.007 × 10⁻⁴ g/100 mL. Dilution factor 100 → stock = 5.007 × 10⁻² g/100 mL = 0.5007 mg mL⁻¹. In 1000 mL: 500.7 mg ≈ 500 mg (100.1% of label).
Safety Precautions & Hazard Control
- Use quartz cuvettes; glass absorbs at 243 nm.
- Degas solvents; bubbles scatter UV light.
- Confirm excipients do not absorb at 243 nm with a placebo blank.
- Work below A = 1.0 for photometric accuracy.
Waste Disposal Protocol
Dilute acidic paracetamol solutions can go to the aqueous drain.
Viva Voce Oral Defense Questions & Answers
Q: What does A(1%, 1 cm) = 715 mean?
A: A solution of 1 g per 100 mL in a 1 cm cell gives absorbance 715; it is the specific absorbance used for direct assay without a full calibration curve.
Q: Why quartz cuvettes here?
A: Glass absorbs strongly below ~320 nm, so at 243 nm only quartz transmits the beam.
Q: What is the main assumption of this direct assay?
A: That nothing else in the tablet (excipients, degradation products) absorbs at 243 nm.