To determine phenol in water by oxidative coupling with 4-aminoantipyrine to a red dye measured at 500 nm.

Level: UniversitySafety: Needs supervisionTime: 60 min
Aim of Experiment: To determine phenol in water by oxidative coupling with 4-aminoantipyrine to a red dye measured at 500 nm.

Theoretical Foundations

Phenols undergo oxidative coupling with 4-aminoantipyrine in the presence of ferricyanide at pH ~10 to give a red antipyrine dye (λmax 500 nm). The reaction requires a free para position on the phenol ring. Absorbance is proportional to phenol from ~0.05 to 5 mg L⁻¹.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV-Vis spectrophotometer
  • Volumetric flasks 100 mL
  • Pipettes
  • Cuvettes
  • pH meter

Chemical Reagents

  • Phenol standard (100 mg L⁻¹)
  • 4-aminoantipyrine solution
  • Potassium ferricyanide solution
  • Ammonium hydroxide buffer (pH 10)
  • Unknown water sample

Step-by-Step Practical Procedure

  1. Pipette 0-5 mL phenol standard into 100 mL flasks (0-5 mg L⁻¹).
  2. Add buffer to pH 10, then 2 mL 4-aminoantipyrine and 2 mL ferricyanide.
  3. Dilute to mark; prepare the unknown identically.
  4. Wait 15 min for full red colour development.
  5. Zero at 500 nm with the reagent blank; read absorbances.
  6. Determine phenol from the calibration curve.

Observations & Expected Results

Red dye forms; A500 linear with phenol; typical slope ≈ 0.30 A·L mg⁻¹.

Worked Stoichiometric Calculations

Problem: Slope is 0.30 A·L mg⁻¹ at 500 nm. A sample gives A = 0.270. Find phenol in mg L⁻¹.

Working: c = 0.270/0.30 = 0.90 mg L⁻¹ phenol.

Safety Precautions & Hazard Control

  • Phenol is toxic and absorbed through skin; wear gloves.
  • Maintain pH 10.0 ± 0.2; coupling fails outside this range.
  • Para-substituted phenols do not couple; the method measures reactive phenols only.
  • Distil the sample first if sulfides or oils interfere.

Waste Disposal Protocol

Collect phenol-containing waste in the toxic organic waste bottle.

Viva Voce Oral Defense Questions & Answers

Q: Why must the para position be free?

A: Oxidative coupling links the antipyrine reagent at the para carbon; a blocked para position prevents dye formation.

Q: Why pH 10?

A: Phenol must be ionized to phenolate for coupling, and ferricyanide oxidation is optimal near pH 10.

Q: What does the method actually measure in a mixed effluent?

A: Only phenols with a free ortho or para position (phenol index), not every phenolic compound.