Phenol Determination: 4-Aminoantipyrine Method at 500 nm
To determine phenol in water by oxidative coupling with 4-aminoantipyrine to a red dye measured at 500 nm.
Theoretical Foundations
Phenols undergo oxidative coupling with 4-aminoantipyrine in the presence of ferricyanide at pH ~10 to give a red antipyrine dye (λmax 500 nm). The reaction requires a free para position on the phenol ring. Absorbance is proportional to phenol from ~0.05 to 5 mg L⁻¹.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks 100 mL
- Pipettes
- Cuvettes
- pH meter
Chemical Reagents
- Phenol standard (100 mg L⁻¹)
- 4-aminoantipyrine solution
- Potassium ferricyanide solution
- Ammonium hydroxide buffer (pH 10)
- Unknown water sample
Step-by-Step Practical Procedure
- Pipette 0-5 mL phenol standard into 100 mL flasks (0-5 mg L⁻¹).
- Add buffer to pH 10, then 2 mL 4-aminoantipyrine and 2 mL ferricyanide.
- Dilute to mark; prepare the unknown identically.
- Wait 15 min for full red colour development.
- Zero at 500 nm with the reagent blank; read absorbances.
- Determine phenol from the calibration curve.
Observations & Expected Results
Red dye forms; A500 linear with phenol; typical slope ≈ 0.30 A·L mg⁻¹.
Worked Stoichiometric Calculations
Problem: Slope is 0.30 A·L mg⁻¹ at 500 nm. A sample gives A = 0.270. Find phenol in mg L⁻¹.
Working: c = 0.270/0.30 = 0.90 mg L⁻¹ phenol.
Safety Precautions & Hazard Control
- Phenol is toxic and absorbed through skin; wear gloves.
- Maintain pH 10.0 ± 0.2; coupling fails outside this range.
- Para-substituted phenols do not couple; the method measures reactive phenols only.
- Distil the sample first if sulfides or oils interfere.
Waste Disposal Protocol
Collect phenol-containing waste in the toxic organic waste bottle.
Viva Voce Oral Defense Questions & Answers
Q: Why must the para position be free?
A: Oxidative coupling links the antipyrine reagent at the para carbon; a blocked para position prevents dye formation.
Q: Why pH 10?
A: Phenol must be ionized to phenolate for coupling, and ferricyanide oxidation is optimal near pH 10.
Q: What does the method actually measure in a mixed effluent?
A: Only phenols with a free ortho or para position (phenol index), not every phenolic compound.