Protein Estimation: Lowry Method at 750 nm
To estimate protein by the Lowry method combining Biuret complexation with Folin-Ciocalteu reduction at 750 nm.
Theoretical Foundations
The Lowry method first forms the Cu-peptide complex (Biuret step), then the Folin-Ciocalteu reagent is reduced by the complex plus tyrosine/tryptophan residues to molybdenum/tungsten blue (λmax 750 nm). The two-step amplification gives sensitivity ~0.01-1 mg mL⁻¹, but timing and pH control are critical.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Test tubes
- Pipettes
- Stopwatch
- Cuvettes
Chemical Reagents
- BSA standard (1 mg mL⁻¹)
- Alkaline copper reagent
- Folin-Ciocalteu reagent (1 N)
- Unknown protein sample
Step-by-Step Practical Procedure
- Pipette 0-1 mL BSA standard into tubes; make to 1 mL (0-1 mg mL⁻¹).
- Add 5 mL alkaline copper reagent; wait 10 min.
- Add 0.5 mL Folin reagent rapidly with immediate vortexing.
- Prepare the unknown identically with exact timing.
- Wait 30 min; zero at 750 nm with blank.
- Read absorbances and determine protein from the curve.
Observations & Expected Results
Blue colour develops; A750 linear with protein; typical slope ≈ 0.90 A·mL mg⁻¹.
Worked Stoichiometric Calculations
Problem: Slope is 0.90 A·mL mg⁻¹ at 750 nm. An unknown gives A = 0.630. Find protein in mg mL⁻¹.
Working: c = 0.630/0.90 = 0.70 mg mL⁻¹.
Safety Precautions & Hazard Control
- Add Folin reagent rapidly with instant mixing; slow addition gives low colour.
- Timing must be identical for standards and unknowns.
- Reducing agents (mercaptoethanol, DTT) cause huge blanks; remove them.
- Folin reagent is acidic and corrosive; wear gloves.
Waste Disposal Protocol
Collect acidic molybdate/tungstate waste in the aqueous heavy-metal waste bottle.
Viva Voce Oral Defense Questions & Answers
Q: What are the two colour-forming steps in Lowry's method?
A: Biuret Cu-peptide complex formation, then reduction of Folin-Ciocalteu reagent to heteropoly blue by the complex and aromatic residues.
Q: Why must Folin reagent be added quickly with vortexing?
A: It is unstable in alkali and decomposes within seconds; only rapid mixing lets it react with the protein complex before decaying.
Q: Which reagents ruin a Lowry assay?
A: Reducing agents like DTT or mercaptoethanol, and also EDTA, Tris and detergents at high levels.