Fluoride Determination: SPADNS Bleaching Method at 570 nm
To determine fluoride in drinking water by bleaching of the red Zr-SPADNS complex, measured at 570 nm.
Theoretical Foundations
The red zirconium-SPADNS complex (λmax 570 nm) is bleached by fluoride, which competes for Zr⁴⁺ forming colourless [ZrF6]²⁻. Absorbance therefore DECREASES as fluoride increases, and the calibration plots absorbance against fluoride with a negative slope. The method covers 0-2 mg L⁻¹ F⁻.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks 50 mL (7)
- Pipettes
- Cuvettes
Chemical Reagents
- NaF standard (10 mg L⁻¹ F⁻)
- SPADNS-Zr reagent (mixed)
- Unknown water sample
Step-by-Step Practical Procedure
- Pipette 0-5 mL fluoride standard into 50 mL flasks (0-1.0 mg L⁻¹).
- Add 10 mL mixed SPADNS-Zr reagent to each; dilute to mark.
- Prepare the unknown identically.
- Wait 10 min; zero at 570 nm with distilled water.
- Read absorbances of standards and sample.
- Plot A versus F⁻ (falling line) and read the unknown.
Observations & Expected Results
Red colour fades progressively with fluoride; calibration is a straight line of negative slope (≈ −0.105 A·L mg⁻¹); blank absorbance ≈ 0.820.
Worked Stoichiometric Calculations
Problem: Blank A = 0.820; sample A = 0.610 at 570 nm; calibration slope = −0.105 A·L mg⁻¹. Find fluoride in mg L⁻¹.
Working: ΔA = 0.820 − 0.610 = 0.210. c = 0.210/0.105 = 2.00 mg L⁻¹ F⁻.
Safety Precautions & Hazard Control
- Aluminium strongly complexes fluoride; remove or mask it.
- High chloride and sulfate cause small positive errors; keep matrix similar.
- Read at a fixed time after mixing; bleaching continues slowly.
- Use fluoride-free distilled water for all dilutions.
Waste Disposal Protocol
Collect zirconium-dye waste in the organic-aqueous waste container.
Viva Voce Oral Defense Questions & Answers
Q: Why does absorbance decrease with more fluoride?
A: Fluoride strips Zr⁴⁺ from the red Zr-SPADNS complex to form colourless [ZrF6]²⁻, so colour is lost as fluoride rises.
Q: How does this calibration differ from normal Beer's law plots?
A: It has a negative slope: the blank gives maximum absorbance and standards fall along a descending line.
Q: Why is aluminium a serious interference?
A: Al³⁺ binds fluoride as [AlF6]³⁻, hiding it from the zirconium reagent and giving low results.