Silica Determination: Molybdenum Yellow Method at 410 nm
To determine dissolved silica in water as the yellow silicomolybdate complex measured at 410 nm.
Theoretical Foundations
Silicic acid reacts with ammonium molybdate in acid to give yellow silicomolybdic acid (λmax ≈ 410 nm). Phosphate forms an analogous complex; it is destroyed selectively with oxalic acid, leaving only the silica colour. Absorbance is proportional to silica up to ~50 mg L⁻¹ SiO2.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks 100 mL (7)
- Pipettes
- Cuvettes
- Plastic bottles for samples
Chemical Reagents
- Sodium silicate standard (100 mg L⁻¹ SiO2)
- Ammonium molybdate (in HCl)
- Oxalic acid solution (10%)
- Unknown water sample
Step-by-Step Practical Procedure
- Pipette 0-10 mL silica standard into 100 mL flasks (0-10 mg L⁻¹).
- Add 2 mL molybdate reagent, mix, wait 10 min.
- Add 2 mL oxalic acid to destroy the phosphomolybdate colour; dilute to mark.
- Prepare the unknown identically.
- Zero at 410 nm with the blank; read absorbances.
- Find silica from the calibration curve.
Observations & Expected Results
Yellow colour develops; oxalic acid bleaches the phosphate contribution; calibration linear at 410 nm with slope ≈ 0.11 A·L mg⁻¹.
Worked Stoichiometric Calculations
Problem: Slope is 0.11 A·L mg⁻¹ at 410 nm. A sample gives A = 0.275. Find SiO2 in mg L⁻¹.
Working: c = 0.275/0.11 = 2.50 mg L⁻¹ SiO2.
Safety Precautions & Hazard Control
- Store samples in plastic, not glass; glass leaches silica.
- Add oxalic acid after the 10 min reaction, not before.
- Keep the acid concentration constant in every flask.
- High iron colours the blank; correct with a sample blank.
Waste Disposal Protocol
Dispose of molybdate solutions in the aqueous heavy-metal waste bottle.
Viva Voce Oral Defense Questions & Answers
Q: Why is oxalic acid added?
A: It selectively destroys the yellow phosphomolybdate complex while leaving silicomolybdate intact, removing phosphate interference.
Q: Why store silica samples in plastic?
A: Glass slowly dissolves, leaching silicate into the sample and giving high results.
Q: Why must timing be consistent?
A: The molybdate reaction rate and the phosphate-masking step are time-dependent; inconsistent timing shifts the calibration.