Phosphate Determination: Molybdenum Blue Method at 880 nm
To determine orthophosphate by the molybdenum blue method with ascorbic acid reduction, measuring at 880 nm.
Theoretical Foundations
Orthophosphate reacts with molybdate in acid to form phosphomolybdic acid, which ascorbic acid reduces to the intensely blue heteropoly complex (λmax ≈ 880 nm). Measuring in the near-IR avoids most coloured interferences. Absorbance is proportional to phosphate over 0-1 mg P L⁻¹.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks 100 mL (7)
- Pipettes
- Water bath (optional)
- Cuvettes
Chemical Reagents
- KH2PO4 standard (10 mg P L⁻¹)
- Ammonium molybdate reagent
- Ascorbic acid solution (fresh)
- Sulfuric acid (2.5 M)
- Unknown phosphate sample
Step-by-Step Practical Procedure
- Pipette 0-8 mL phosphate standard into 100 mL flasks (0-0.8 mg P L⁻¹).
- Add 8 mL molybdate reagent and 2 mL ascorbic acid to each; dilute to mark.
- Prepare the unknown identically.
- Allow 20 min at room temperature for full blue colour.
- Zero at 880 nm with the reagent blank; read absorbances.
- Construct the calibration curve and evaluate the unknown.
Observations & Expected Results
Deep-blue colour develops gradually; absorbance at 880 nm is linear with phosphate; slope typically ≈ 0.73 A·L mg⁻¹.
Worked Stoichiometric Calculations
Problem: Calibration slope is 0.7275 A·L mg⁻¹ at 880 nm. A sample gives A = 0.398. Find mg P L⁻¹.
Working: c = 0.398/0.7275 = 0.547 mg P L⁻¹.
Safety Precautions & Hazard Control
- Use freshly prepared ascorbic acid; old solutions reduce unevenly.
- Silicate also forms a blue complex; control reaction time to minimize it.
- Arsenate interferes identically; remove or correct if present.
- Keep acid strength constant; colour intensity depends on acidity.
Waste Disposal Protocol
Dispose of molybdate-containing solutions in the aqueous heavy-metal waste bottle.
Viva Voce Oral Defense Questions & Answers
Q: Why is 880 nm preferred over the visible region?
A: The molybdenum blue peak at 880 nm lies away from most natural water colours, reducing background interference.
Q: Why must ascorbic acid be fresh?
A: Oxidised ascorbic acid reduces molybdate inconsistently, giving non-reproducible colour development.
Q: Which common ion mimics phosphate here?
A: Silicate forms a similar heteropoly blue; arsenate also responds, so both are potential interferences.