CHEMVEN

Chromium(VI) Determination: Diphenylcarbazide Method at 540 nm

To determine trace Cr(VI) in water by the red-violet diphenylcarbazide complex at 540 nm.

Level: UniversitySafety: Needs supervisionTime: 90 min
Aim of Experiment: To determine trace Cr(VI) in water by the red-violet diphenylcarbazide complex at 540 nm.

Theoretical Foundations

Diphenylcarbazide is oxidised by Cr(VI) in acid to diphenylcarbazone, which chelates the reduced Cr(III) giving a red-violet complex (λmax 540 nm, ε ≈ 4.2 × 10⁴ L mol⁻¹ cm⁻¹). The very high absorptivity allows determination down to ~0.01 mg L⁻¹. The reaction is specific for Cr(VI); Cr(III) does not react.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV-Vis spectrophotometer
  • Volumetric flasks 100 mL (7)
  • Pipettes
  • Cuvettes
  • pH meter or paper

Chemical Reagents

  • K2Cr2O7 standard (10 mg L⁻¹ as Cr)
  • Diphenylcarbazide (0.25% in acetone)
  • Sulfuric acid (1 M)
  • Unknown water sample

Step-by-Step Practical Procedure

  1. Pipette 0-5 mL Cr(VI) standard into 100 mL flasks (0-0.5 mg L⁻¹).
  2. Add 2 mL 1 M H2SO4 and 2 mL diphenylcarbazide to each; dilute to mark.
  3. Prepare the unknown identically.
  4. Allow 10 min for full colour development.
  5. Zero at 540 nm with the reagent blank; read absorbances.
  6. Determine the unknown from the calibration curve.

Observations & Expected Results

Red-violet colour develops within minutes; calibration linear to ~0.5 mg L⁻¹ with steep slope reflecting the high ε at 540 nm.

Worked Stoichiometric Calculations

Problem: An effluent sample gives A = 0.685 at 540 nm (ε = 4.20 × 10⁴, l = 1 cm). Find Cr(VI) in mg L⁻¹.

Working: c = 0.685/4.20 × 10⁴ = 1.63 × 10⁻⁵ mol L⁻¹. mg L⁻¹ = 1.63 × 10⁻⁵ × 52.00 × 1000 = 0.848 mg L⁻¹.

Safety Precautions & Hazard Control

  • Cr(VI) standards are carcinogenic; wear gloves and avoid ingestion.
  • Develop colour in acid medium; the reaction fails at neutral pH.
  • Read within 30 min; the complex slowly fades.
  • Oxidize any Cr(III) to Cr(VI) first if total chromium is required.

Waste Disposal Protocol

Collect all Cr(VI)-containing solutions in the chromium waste bottle for professional disposal.

Viva Voce Oral Defense Questions & Answers

Q: Why is this method specific for Cr(VI)?

A: Only Cr(VI) oxidises diphenylcarbazide to the carbazone that chelates chromium; Cr(III) cannot perform this oxidation.

Q: Why is the method so sensitive?

A: The complex has an exceptionally high molar absorptivity (~4.2 × 10⁴ L mol⁻¹ cm⁻¹), so tiny concentrations give measurable absorbance.

Q: What happens if the solution is not acidic?

A: The redox reaction between Cr(VI) and diphenylcarbazide requires H⁺; in neutral solution little colour develops.

← Back to Experiments Directory