Iron Determination: 1,10-Phenanthroline Method at 510 nm
To determine total iron in water by reducing Fe(III) to Fe(II) and forming the orange-red [Fe(phen)3]²⁺ complex measured at 510 nm.
Theoretical Foundations
1,10-phenanthroline forms the stable orange-red complex [Fe(phen)3]²⁺ with Fe(II), λmax = 510 nm, ε = 1.11 × 10⁴ L mol⁻¹ cm⁻¹. Hydroxylamine hydrochloride reduces all iron to Fe(II), and acetate buffer holds pH 3-5 where the complex is stable. The colour is stable for hours, unlike the thiocyanate method.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks 100 mL (7)
- Pipettes
- Cuvettes
- pH paper
Chemical Reagents
- Fe standard solution (10 mg L⁻¹)
- Hydroxylamine hydrochloride (10% w/v)
- Sodium acetate buffer (pH 4.5)
- 1,10-phenanthroline solution (0.25% w/v)
- Unknown water sample
Step-by-Step Practical Procedure
- Add 0-8 mL Fe standard to 100 mL flasks (0-0.8 mg L⁻¹).
- Add 2 mL hydroxylamine solution and 10 mL acetate buffer to each.
- Add 10 mL phenanthroline solution; dilute to mark and mix.
- Treat the unknown identically using a 5.00 mL aliquot.
- Wait 15 min for full colour development.
- Zero at 510 nm with the blank and read all absorbances.
Observations & Expected Results
Orange-red colour develops, stable for hours; calibration linear with slope corresponding to ε = 1.11 × 10⁴ L mol⁻¹ cm⁻¹ at 510 nm.
Worked Stoichiometric Calculations
Problem: A 5.00 mL water sample diluted to 100 mL gives A = 0.512 at 510 nm (ε = 1.11 × 10⁴, l = 1 cm). Find Fe in the original sample in mg L⁻¹.
Working: c(flask) = 0.512/1.11 × 10⁴ = 4.61 × 10⁻⁵ mol L⁻¹. Mass in 0.100 L = 4.61 × 10⁻⁶ × 55.85 = 2.58 × 10⁻⁴ g = 0.258 mg. Original conc. = 0.258 mg/0.005 L = 51.5 mg L⁻¹.
Safety Precautions & Hazard Control
- Ensure complete reduction with hydroxylamine before adding phenanthroline.
- Maintain pH 3-5 with buffer; outside this range the complex does not form fully.
- Allow full 15 min for colour development.
- Avoid phosphate and strong complexing agents that compete for Fe.
Waste Disposal Protocol
Collect phenanthroline-containing waste in the organic-aqueous waste bottle.
Viva Voce Oral Defense Questions & Answers
Q: What is the role of hydroxylamine hydrochloride?
A: It reduces Fe(III) to Fe(II), because phenanthroline forms the coloured complex only with Fe(II).
Q: Why is this method preferred over thiocyanate for trace iron?
A: The [Fe(phen)3]²⁺ colour is far more stable in light and the molar absorptivity is high, giving better precision at low levels.
Q: What pH range is required and why?
A: pH 3-5: below this complex formation is incomplete, above it Fe(II) hydrolyses and precipitates.