Iron Determination: Thiocyanate Colorimetric Method at 480 nm
To determine iron(III) in a sample by forming the blood-red Fe(SCN)²⁺ complex and measuring absorbance at 480 nm against a calibration curve.
Theoretical Foundations
Fe³⁺ reacts with excess SCN⁻ to give the intensely red complex [Fe(SCN)]²⁺ with λmax ≈ 480 nm and ε ≈ 1.1 × 10⁴ L mol⁻¹ cm⁻¹. With thiocyanate in large excess, complex formation is essentially complete and absorbance is proportional to total iron. The colour fades slowly in light, so readings are taken promptly.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Volumetric flasks 50 mL (6)
- Pipettes
- Cuvettes
- Stopwatch
Chemical Reagents
- Fe(III) standard solution (100 mg L⁻¹)
- Potassium thiocyanate solution (0.5 M)
- Dilute nitric acid (1 M)
- Distilled water
- Unknown iron sample
Step-by-Step Practical Procedure
- Pipette 0, 1, 2, 3, 4, 5 mL of Fe standard into 50 mL flasks (0-10 mg L⁻¹).
- Add 5 mL dilute HNO3 and 5 mL KSCN to each flask; dilute to mark.
- Prepare the unknown the same way using a suitable aliquot.
- Zero the instrument at 480 nm with the reagent blank.
- Measure absorbances within 15 min of colour development.
- Plot calibration curve and read the unknown concentration.
Observations & Expected Results
Standards develop a blood-red colour deepening with concentration; calibration is linear up to ~10 mg L⁻¹ with slope ≈ 0.197 A·L mg⁻¹ (ε ≈ 1.1 × 10⁴).
Worked Stoichiometric Calculations
Problem: An unknown gives A = 0.620 at 480 nm (l = 1 cm, ε = 1.10 × 10⁴ L mol⁻¹ cm⁻¹). Find Fe concentration in mol L⁻¹ and mg L⁻¹.
Working: c = A/(εl) = 0.620/1.10 × 10⁴ = 5.64 × 10⁻⁵ mol L⁻¹. In mg L⁻¹: 5.64 × 10⁻⁵ × 55.85 × 1000 = 3.15 mg L⁻¹.
Safety Precautions & Hazard Control
- Add thiocyanate in large excess to drive complex formation to completion.
- Measure promptly; the red colour fades on standing in light.
- Keep acidity constant in all flasks; hydrolysis of Fe³⁺ alters the colour.
- Oxidize any Fe(II) to Fe(III) before analysis since Fe(II) gives no colour.
Waste Disposal Protocol
Collect thiocyanate/iron waste in the aqueous heavy-metal waste container.
Viva Voce Oral Defense Questions & Answers
Q: Why is SCN⁻ added in large excess?
A: To shift the equilibrium Fe³⁺ + SCN⁻ ⇌ [Fe(SCN)]²⁺ fully to the right so absorbance measures total iron.
Q: Why does the method require Fe(III), not Fe(II)?
A: Only Fe(III) forms the coloured thiocyanate complex; Fe(II) must first be oxidized, e.g. with nitric acid.
Q: Why measure quickly after adding reagent?
A: The complex is photochemically reduced over time, causing the red colour and absorbance to fade.