Verification of Beer-Lambert Law: Absorbance of Potassium Permanganate
To verify that absorbance is directly proportional to concentration (A = εcl) using standard KMnO4 solutions at 525 nm.
Theoretical Foundations
The Beer-Lambert law states that absorbance A = εcl, where ε is molar absorptivity, c is concentration and l is path length. For a fixed path length, a plot of A versus c must be a straight line through the origin. Permanganate ion has an intense purple charge-transfer band with λmax near 525 nm, making it ideal for this verification.
Required Apparatus & Chemicals
Laboratory Apparatus
- UV-Vis spectrophotometer
- Glass cuvettes (1 cm)
- Volumetric flasks 100 mL (5)
- Pipettes 1, 2, 5 mL
- Beakers
- Wash bottle
Chemical Reagents
- KMnO4 stock solution (1.0 × 10⁻³ M)
- KMnO4 (AR grade, for stock preparation)
- Distilled water (blank and diluent)
Step-by-Step Practical Procedure
- Prepare standards of 0.5, 1.0, 2.0, 3.0 and 4.0 × 10⁻⁴ M KMnO4 by diluting the stock in 100 mL flasks.
- Switch on the spectrophotometer and allow 15 min warm-up.
- Set wavelength to 525 nm and zero absorbance with distilled water blank.
- Measure absorbance of each standard in order of increasing concentration.
- Rinse cuvette with the next solution before each reading.
- Plot absorbance versus concentration and determine slope and intercept.
Observations & Expected Results
Absorbance rises linearly with concentration (e.g. 0.12, 0.24, 0.48, 0.72, 0.96); the calibration line passes through the origin with slope ≈ εl ≈ 2.40 × 10³ L mol⁻¹.
Worked Stoichiometric Calculations
Problem: A KMnO4 solution of concentration 2.0 × 10⁻⁴ mol L⁻¹ gives what absorbance at 525 nm if ε = 2.40 × 10³ L mol⁻¹ cm⁻¹ and l = 1 cm?
Working: A = εcl = (2.40 × 10³)(2.0 × 10⁻⁴)(1) = 0.480. The measured value should match within experimental error, confirming the law.
Safety Precautions & Hazard Control
- Use freshly prepared dilute KMnO4; it slowly decomposes in light.
- Handle cuvettes by the frosted sides only.
- Zero the instrument with the blank before every set of readings.
- Work in the 0.1-1.0 absorbance range for best accuracy.
Waste Disposal Protocol
Collect dilute permanganate waste in the aqueous oxidizer waste bottle; do not pour concentrated solutions down the drain.
Viva Voce Oral Defense Questions & Answers
Q: Why is absorbance measured at λmax (525 nm)?
A: Sensitivity is highest there and small wavelength errors cause the least change in absorbance, since dA/dλ ≈ 0 at the peak.
Q: Why must the blank be the pure solvent?
A: To cancel absorbance from the solvent, cuvette walls and reflections, so the reading reflects only the analyte.
Q: Name two causes of deviation from Beer's law.
A: Stray light and high concentrations causing molecular association; chemical deviations from equilibria such as dimerization or protonation shifts.