Verification of Beer-Lambert Law: Absorbance of Potassium Permanganate

To verify that absorbance is directly proportional to concentration (A = εcl) using standard KMnO4 solutions at 525 nm.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To verify that absorbance is directly proportional to concentration (A = εcl) using standard KMnO4 solutions at 525 nm.

Theoretical Foundations

The Beer-Lambert law states that absorbance A = εcl, where ε is molar absorptivity, c is concentration and l is path length. For a fixed path length, a plot of A versus c must be a straight line through the origin. Permanganate ion has an intense purple charge-transfer band with λmax near 525 nm, making it ideal for this verification.

Required Apparatus & Chemicals

Laboratory Apparatus

  • UV-Vis spectrophotometer
  • Glass cuvettes (1 cm)
  • Volumetric flasks 100 mL (5)
  • Pipettes 1, 2, 5 mL
  • Beakers
  • Wash bottle

Chemical Reagents

  • KMnO4 stock solution (1.0 × 10⁻³ M)
  • KMnO4 (AR grade, for stock preparation)
  • Distilled water (blank and diluent)

Step-by-Step Practical Procedure

  1. Prepare standards of 0.5, 1.0, 2.0, 3.0 and 4.0 × 10⁻⁴ M KMnO4 by diluting the stock in 100 mL flasks.
  2. Switch on the spectrophotometer and allow 15 min warm-up.
  3. Set wavelength to 525 nm and zero absorbance with distilled water blank.
  4. Measure absorbance of each standard in order of increasing concentration.
  5. Rinse cuvette with the next solution before each reading.
  6. Plot absorbance versus concentration and determine slope and intercept.

Observations & Expected Results

Absorbance rises linearly with concentration (e.g. 0.12, 0.24, 0.48, 0.72, 0.96); the calibration line passes through the origin with slope ≈ εl ≈ 2.40 × 10³ L mol⁻¹.

Worked Stoichiometric Calculations

Problem: A KMnO4 solution of concentration 2.0 × 10⁻⁴ mol L⁻¹ gives what absorbance at 525 nm if ε = 2.40 × 10³ L mol⁻¹ cm⁻¹ and l = 1 cm?

Working: A = εcl = (2.40 × 10³)(2.0 × 10⁻⁴)(1) = 0.480. The measured value should match within experimental error, confirming the law.

Safety Precautions & Hazard Control

  • Use freshly prepared dilute KMnO4; it slowly decomposes in light.
  • Handle cuvettes by the frosted sides only.
  • Zero the instrument with the blank before every set of readings.
  • Work in the 0.1-1.0 absorbance range for best accuracy.

Waste Disposal Protocol

Collect dilute permanganate waste in the aqueous oxidizer waste bottle; do not pour concentrated solutions down the drain.

Viva Voce Oral Defense Questions & Answers

Q: Why is absorbance measured at λmax (525 nm)?

A: Sensitivity is highest there and small wavelength errors cause the least change in absorbance, since dA/dλ ≈ 0 at the peak.

Q: Why must the blank be the pure solvent?

A: To cancel absorbance from the solvent, cuvette walls and reflections, so the reading reflects only the analyte.

Q: Name two causes of deviation from Beer's law.

A: Stray light and high concentrations causing molecular association; chemical deviations from equilibria such as dimerization or protonation shifts.