Thermogravimetry of Calcium Oxalate Monohydrate: Three-Step Decomposition
Record the stepwise mass-loss curve of CaC2O4·H2O and identify each plateau compound.
Theoretical Foundations
Calcium oxalate monohydrate is the classic TGA standard because it decomposes in three well-separated steps: CaC2O4·H2O → CaC2O4 + H2O (12.33% loss), CaC2O4 → CaCO3 + CO (19.17% loss), CaCO3 → CaO + CO2 (30.12% loss). Each plateau corresponds to a stoichiometric weighing form, linking thermogravimetry directly to gravimetric analysis. A manual stepwise ignition in a furnace reproduces the TGA curve.
Required Apparatus & Chemicals
Laboratory Apparatus
- Analytical balance
- Porcelain crucible (weighed)
- Muffle furnace
- Desiccator
- Tongs
- Thermometer/pyrometer
Chemical Reagents
- Calcium oxalate monohydrate, pure
- Silica gel desiccant (for desiccator)
- Distilled water (glassware rinsing)
Step-by-Step Practical Procedure
- Weigh accurately 2.000 g CaC2O4·H2O into a weighed crucible.
- Heat at 200 C for 1 hour; cool in a desiccator and weigh (plateau 1).
- Heat at 500 C for 1 hour; cool and weigh (plateau 2).
- Heat at 900 C for 1 hour; cool and weigh (plateau 3).
- Calculate the mass at each plateau and the % loss of each step.
- Compare with the theoretical plateau masses.
Observations & Expected Results
White solid throughout; mass drops in three distinct steps with flat plateaus between.
Worked Stoichiometric Calculations
Problem: 2.000 g CaC2O4·H2O (M = 146.11) heated stepwise. Calculate the theoretical mass at each plateau: CaC2O4 (128.10), CaCO3 (100.09), CaO (56.08).
Working: CaC2O4: 2.000 × 128.10/146.11 = 1.753 g. CaCO3: 2.000 × 100.09/146.11 = 1.370 g. CaO: 2.000 × 56.08/146.11 = 0.7676 g ≈ 0.768 g. Step losses: 12.33%, 19.17%, 30.12%.
Safety Precautions & Hazard Control
- Hold each temperature long enough to reach the plateau.
- Cool in a desiccator for a fixed time before each weighing.
- Vent CO safely; the furnace area must be ventilated.
Waste Disposal Protocol
Residues to solid waste.
Viva Voce Oral Defense Questions & Answers
Q: Why is calcium oxalate monohydrate a TGA calibration standard?
A: Its three decomposition steps are sharp, well separated in temperature and exactly stoichiometric, so they check both the mass and temperature calibration of the instrument.
Q: What is a TGA plateau?
A: A temperature range over which the mass is constant, indicating a thermally stable (weighable) compound.