Copper in Electroplating Solution by Electrogravimetry
Determine the copper concentration (g/L) of an acid copper plating bath.
Theoretical Foundations
Plating baths are assayed to maintain metal concentration. An aliquot is electrolysed directly: Cu2+ + 2e- → Cu(s). The cathode gain divided by the aliquot volume gives g/L Cu. Acid copper baths run typically 20-80 g/L Cu; the result guides replenishment additions.
Required Apparatus & Chemicals
Laboratory Apparatus
- 25 mL pipette
- Analytical balance
- Platinum electrodes
- DC power supply
- 250 mL beaker
- Oven
- Desiccator
Chemical Reagents
- Acid copper plating bath sample
- Sulfuric acid, dilute (if needed)
- Distilled water
- Ethanol
Step-by-Step Practical Procedure
- Pipette 25.00 mL of bath into a beaker; dilute to 150 mL if very concentrated.
- Weigh the Pt cathode.
- Electrolyse at 2 A with stirring for 45 minutes.
- Test completeness; wash with current on; rinse with ethanol and dry.
- Weigh the cathode; calculate g/L Cu.
Observations & Expected Results
Blue bath fades as copper deposits; typical gain about 1.9 g from 25 mL of bath.
Worked Stoichiometric Calculations
Problem: 25.00 mL of bath gave a cathode gain of 1.875 g Cu. Calculate the bath concentration in g/L.
Working: Concentration = 1.875 g / 0.02500 L = 75.0 g/L Cu.
Safety Precautions & Hazard Control
- Pipette viscous bath samples carefully and rinse the pipette into the cell.
- Ensure complete deposition; test on a fresh cathode surface.
- Keep current on during washing.
Waste Disposal Protocol
Depleted bath sample to heavy-metal waste.
Viva Voce Oral Defense Questions & Answers
Q: Why monitor plating bath copper concentration?
A: It controls the plating rate and deposit quality; depletion causes slow, poor plating, so the assay triggers replenishment.
Q: Why is this method well suited to plating baths?
A: The bath already contains only the analyte in a conductive acid medium — the sample can be electrolysed almost directly, with no separation needed.