Silica in Portland Cement by Dehydration
Determine SiO2 in Portland cement by double dehydration of silicic acid.
Theoretical Foundations
Acid decomposition of cement gives gelatinous silicic acid: SiO3 2- + 2H+ → H2SiO3. Evaporating to dryness and baking dehydrates it: H2SiO3 → SiO2 + H2O. Silica is filtered off, ignited and weighed directly — no gravimetric factor is needed since the weighing form is the analyte oxide. A second dehydration recovers the few percent of silica that escapes the first. Typical cement is ~20% SiO2.
Required Apparatus & Chemicals
Laboratory Apparatus
- Analytical balance
- 250 mL beaker
- Hot plate
- Oven at 110 C and muffle furnace
- Ashless filter paper
- Platinum crucible
- Desiccator
- Hydrofluoric acid handling kit
Chemical Reagents
- Portland cement sample
- Hydrochloric acid, concentrated
- Hydrofluoric acid (for impurity correction)
- Sulfuric acid, dilute
- Distilled water
Step-by-Step Practical Procedure
- Weigh accurately 1.0 g cement into a beaker.
- Add 15 mL conc. HCl; warm until decomposed.
- Evaporate to dryness on a hot plate; bake 1 hour at 110 C.
- Moisten with conc. HCl, add 50 mL hot water, digest and filter through ashless paper.
- Evaporate the filtrate again to dryness, bake and filter (second dehydration); combine precipitates.
- Ignite the combined silica at 1000 C to constant mass; weigh.
- Treat with HF + H2SO4, evaporate, reignite and reweigh; the mass loss is true SiO2.
- Calculate % SiO2.
Observations & Expected Results
Gelatinous silica on the filter; white ignited SiO2, typically about 0.21 g from 1 g cement.
Worked Stoichiometric Calculations
Problem: 1.000 g of cement gave 0.2135 g of HF-volatile SiO2. Calculate % SiO2.
Working: % SiO2 = (0.2135/1.000) × 100 = 21.35%, a normal Portland cement value.
Safety Precautions & Hazard Control
- Bake the dehydration fully; incomplete baking leaves soluble silica.
- HF is extremely hazardous: full PPE, fume cupboard and calcium gluconate gel available.
- Ignite to constant mass; silica adsorbs water strongly.
Waste Disposal Protocol
HF-containing residues to dedicated fluoride waste; never mix with glass containers.
Viva Voce Oral Defense Questions & Answers
Q: Why dehydrate twice?
A: A few percent of silicic acid stays colloidal through the first filtration; the second evaporation recovers it, ensuring quantitative recovery.
Q: What is the HF treatment for?
A: HF volatilises SiO2 as SiF4; non-volatile impurities (Al2O3, Fe2O3) remain, so the mass loss gives true SiO2 free of contaminants.