Potassium in Fruit Juice by K+ Ion-Selective Electrode
To determine potassium in orange juice with a K+ ISE using the standard-addition method.
Theoretical Foundations
The valinomycin-based K+ electrode follows E = const + 0.05916 log[K+] at 25 C, with good selectivity over Na+. Fruit juice is a complex matrix (acids, sugars), so standard addition is used: the potential shift on spiking with known K+ gives the original concentration, cancelling matrix effects on the slope and junction.
Required Apparatus & Chemicals
Laboratory Apparatus
- Potassium ISE + reference
- mV meter
- Beakers
- Micropipette
- Stirrer
Chemical Reagents
- KCl standard (1000 mg/L K+)
- Ionic strength adjuster
- Orange juice (filtered)
- Distilled water
Step-by-Step Practical Procedure
- Filter the juice and add ISA.
- Measure E1 of 50.0 mL juice.
- Spike with 0.50 mL of 1000 mg/L K+ standard; record E2.
- Check the electrode slope with two KCl standards.
- Calculate [K+] from the potential shift.
- Express as mg K+ per 100 mL juice.
Observations & Expected Results
Orange juice typically reads 150-250 mg K+ per 100 mL.
Worked Stoichiometric Calculations
Problem: E1 = 96 mV; after spiking 0.50 mL of 1000 mg/L into 50.0 mL, E2 = 104 mV; S = +57 mV/decade. Find [K+] in mg/L.
Working: dE = +8 mV. C_x = (C_s V_s/V_x)/(10^(dE/S) - 1) = (1000 x 0.5/50)/(10^(8/57) - 1) = 10/(1.381 - 1) = 26.2 mg/L. (Low reading suggests a diluted sample; real juice is ~2000 mg/L.)
Safety Precautions & Hazard Control
- Filter out pulp that could foul the membrane.
- Add ISA uniformly.
- Condition the electrode before low-level work.
Waste Disposal Protocol
Juice and standards to drain.
Viva Voce Oral Defense Questions & Answers
Q: Why is valinomycin used in the membrane?
A: It is a neutral carrier that selectively complexes K+ over Na+ (~10^4 selectivity), giving the electrode its specificity.
Q: Why standard addition for juice?
A: Sugars, acids and varying ionic strength shift the response; spiking the sample itself cancels these matrix effects.