CHEMVEN

Capillary Electrophoresis Concept: Efficiency and Plate Numbers (Dry-Lab)

To calculate separation efficiency (theoretical plates) from a capillary electrophoresis electropherogram as a dry-lab exercise.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To calculate separation efficiency (theoretical plates) from a capillary electrophoresis electropherogram as a dry-lab exercise.

Theoretical Foundations

In CE, electroosmotic flow sweeps all species toward the detector while electrophoretic mobility separates them. Efficiency is very high (10^5-10^6 plates) because there is no packing and the flat flow profile minimises broadening. Plate number N = 16 (tR/W)^2 (or 5.54 (tR/W1/2)^2); resolution Rs = 2(t2-t1)/(W1+W2).

Required Apparatus & Chemicals

Laboratory Apparatus

  • (Dry-lab) Electropherogram with migration times and peak widths
  • Ruler
  • Calculator
  • High-voltage CE power supply (concept)
  • Fused-silica capillary cartridge (concept)

Chemical Reagents

  • (Dry-lab) Fused-silica capillary
  • Background electrolyte
  • Analyte mixture

Step-by-Step Practical Procedure

  1. Measure migration time tR and baseline width W for each peak.
  2. Calculate N for each peak using N = 16(tR/W)^2.
  3. Calculate resolution between adjacent peaks.
  4. Compare N with typical HPLC values (~10^4).
  5. Explain why CE gives such high plate numbers.
  6. State the migration order for cations, neutrals and anions.

Observations & Expected Results

N values of 100,000-300,000 are typical - an order of magnitude above HPLC.

Worked Stoichiometric Calculations

Problem: Peak: tR = 240 s, W = 6.0 s. Calculate N.

Working: N = 16 x (240/6.0)^2 = 16 x 1600 = 25,600 plates (for a short capillary; longer capillaries give far more).

Safety Precautions & Hazard Control

  • Dry-lab only; real CE uses high voltage (up to 30 kV) - interlocked instruments only.
  • Never look into an open high-voltage compartment.
  • Capillaries are fragile; handle with care.

Waste Disposal Protocol

No wet waste in the dry-lab version.

Viva Voce Oral Defense Questions & Answers

Q: Why is CE more efficient than HPLC?

A: No stationary-phase mass transfer and a flat electroosmotic flow profile give almost no band broadening, so plate numbers reach 10^5-10^6.

Q: State the migration order and why.

A: Cations first (electrophoresis + EOF same direction), neutrals with the EOF, anions last (electrophoresis opposes EOF but EOF usually wins).

← Back to Experiments Directory