Gel Electrophoresis Concept: DNA Fragment Sizing (Dry-Lab)
To size unknown DNA fragments from an agarose gel image using a molecular-weight ladder as a dry-lab exercise.
Theoretical Foundations
In agarose gel electrophoresis, DNA (uniform negative charge per base pair) migrates toward the anode; the gel sieves by size so mobility decreases with fragment length. A plot of log(base pairs) versus migration distance for the ladder is linear; unknown bands are sized by interpolation. Ethidium bromide or safer dyes visualise bands under UV.
Required Apparatus & Chemicals
Laboratory Apparatus
- (Dry-lab) Gel photograph with ladder and samples
- Ruler
- UV transilluminator with shield (concept)
- Gel documentation camera (concept)
- Micropipettes (concept)
Chemical Reagents
- (Dry-lab) Agarose
- TAE buffer
- DNA ladder (100 bp - 3000 bp)
- Loading dye
Step-by-Step Practical Procedure
- Measure migration distances of all ladder bands from the wells.
- Plot log(bp) versus distance for the ladder.
- Fit a straight line (excluding the extreme bands if they deviate).
- Measure the unknown bands' distances.
- Read their sizes from the calibration line.
- Estimate the mass of DNA if a quantitative ladder was used.
Observations & Expected Results
The ladder plot is linear (r > 0.99); unknowns typically size to within 5-10% of their true length.
Worked Stoichiometric Calculations
Problem: Ladder line: log(bp) = 3.90 - 0.080 x d (d in mm). An unknown band migrated 22 mm. What is its size?
Working: log(bp) = 3.90 - 0.080 x 22 = 2.14; bp = 10^2.14 = 138 bp (~140 bp).
Safety Precautions & Hazard Control
- Dry-lab only; real ethidium bromide is mutagenic - use safer dyes and gloves in practice.
- Never look at UV transilluminators without a shield.
- Load wells carefully to avoid cross-contamination.
Waste Disposal Protocol
No wet waste in the dry-lab version; real gels with stain go to biohazard/chemical waste.
Viva Voce Oral Defense Questions & Answers
Q: Why does DNA migrate by size rather than charge?
A: Each base pair carries the same charge, so charge-to-mass is constant; the gel matrix then separates purely by sieving (size).
Q: Why plot log(bp) and not bp?
A: Mobility is approximately linear with log(molecular weight) over the resolving range of the gel.
Q: What causes band smiling?
A: Uneven heating or buffer depletion: the centre runs hotter/faster than the edges.