Gel Electrophoresis Concept: DNA Fragment Sizing (Dry-Lab)

To size unknown DNA fragments from an agarose gel image using a molecular-weight ladder as a dry-lab exercise.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To size unknown DNA fragments from an agarose gel image using a molecular-weight ladder as a dry-lab exercise.

Theoretical Foundations

In agarose gel electrophoresis, DNA (uniform negative charge per base pair) migrates toward the anode; the gel sieves by size so mobility decreases with fragment length. A plot of log(base pairs) versus migration distance for the ladder is linear; unknown bands are sized by interpolation. Ethidium bromide or safer dyes visualise bands under UV.

Required Apparatus & Chemicals

Laboratory Apparatus

  • (Dry-lab) Gel photograph with ladder and samples
  • Ruler
  • UV transilluminator with shield (concept)
  • Gel documentation camera (concept)
  • Micropipettes (concept)

Chemical Reagents

  • (Dry-lab) Agarose
  • TAE buffer
  • DNA ladder (100 bp - 3000 bp)
  • Loading dye

Step-by-Step Practical Procedure

  1. Measure migration distances of all ladder bands from the wells.
  2. Plot log(bp) versus distance for the ladder.
  3. Fit a straight line (excluding the extreme bands if they deviate).
  4. Measure the unknown bands' distances.
  5. Read their sizes from the calibration line.
  6. Estimate the mass of DNA if a quantitative ladder was used.

Observations & Expected Results

The ladder plot is linear (r > 0.99); unknowns typically size to within 5-10% of their true length.

Worked Stoichiometric Calculations

Problem: Ladder line: log(bp) = 3.90 - 0.080 x d (d in mm). An unknown band migrated 22 mm. What is its size?

Working: log(bp) = 3.90 - 0.080 x 22 = 2.14; bp = 10^2.14 = 138 bp (~140 bp).

Safety Precautions & Hazard Control

  • Dry-lab only; real ethidium bromide is mutagenic - use safer dyes and gloves in practice.
  • Never look at UV transilluminators without a shield.
  • Load wells carefully to avoid cross-contamination.

Waste Disposal Protocol

No wet waste in the dry-lab version; real gels with stain go to biohazard/chemical waste.

Viva Voce Oral Defense Questions & Answers

Q: Why does DNA migrate by size rather than charge?

A: Each base pair carries the same charge, so charge-to-mass is constant; the gel matrix then separates purely by sieving (size).

Q: Why plot log(bp) and not bp?

A: Mobility is approximately linear with log(molecular weight) over the resolving range of the gel.

Q: What causes band smiling?

A: Uneven heating or buffer depletion: the centre runs hotter/faster than the edges.