Anodic Stripping Concept: Trace Lead in Water (Dry-Lab)

To understand anodic stripping voltammetry for trace Pb2+ as a dry-lab exercise with standard-addition quantitation.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To understand anodic stripping voltammetry for trace Pb2+ as a dry-lab exercise with standard-addition quantitation.

Theoretical Foundations

ASV has two steps: (1) deposition - Pb2+ is reduced into a mercury-film electrode at -0.8 V with stirring, preconcentrating it; (2) stripping - the potential is scanned anodically and Pb oxidises back, giving a sharp peak whose area/height is proportional to concentration. Preconcentration gives detection limits down to ~10^-10 M. Quantitation is by standard addition.

Required Apparatus & Chemicals

Laboratory Apparatus

  • (Dry-lab) Stripping voltammograms: sample and two standard additions
  • Ruler
  • Mercury-film electrode (concept)
  • Deposition timer (concept)
  • Magnetic stirrer (concept)

Chemical Reagents

  • (Dry-lab) Water sample
  • Pb2+ standard
  • Acetate buffer
  • Mercury film (concept)

Step-by-Step Practical Procedure

  1. Measure the Pb stripping peak height for the sample.
  2. Measure peak heights after each standard addition.
  3. Plot peak height versus added concentration.
  4. Extrapolate to zero signal to find the original concentration.
  5. Convert to ug/L (ppb) and compare with the 10 ppb WHO guideline.
  6. Explain the role of the deposition step.

Observations & Expected Results

Peak heights grow linearly with added Pb; extrapolation typically gives low-ppb levels for tap water.

Worked Stoichiometric Calculations

Problem: Sample peak = 12 nA; +10 ppb spike gives 32 nA. Find the sample concentration.

Working: Slope = (32-12)/10 = 2.0 nA/ppb. C = 12/2.0 = 6.0 ppb - below the 10 ppb WHO guideline.

Safety Precautions & Hazard Control

  • Dry-lab only: real ASV uses mercury films handled by trained staff.
  • In practice, control deposition time precisely; it sets sensitivity.
  • Deaerate the cell to remove O2 interference.

Waste Disposal Protocol

No wet waste in the dry-lab version; real mercury waste to dedicated collection.

Viva Voce Oral Defense Questions & Answers

Q: Why is ASV so sensitive?

A: The deposition step preconcentrates the analyte from a large solution volume into a tiny electrode volume, amplifying the stripping signal by 100-1000x.

Q: Why standard addition?

A: It corrects for matrix effects on deposition efficiency in real water samples.

Q: What sets the deposition potential?

A: It must be negative enough to reduce Pb2+ (on the diffusion plateau) but not so negative that other ions or H+ interfere.