Anodic Stripping Concept: Trace Lead in Water (Dry-Lab)
To understand anodic stripping voltammetry for trace Pb2+ as a dry-lab exercise with standard-addition quantitation.
Theoretical Foundations
ASV has two steps: (1) deposition - Pb2+ is reduced into a mercury-film electrode at -0.8 V with stirring, preconcentrating it; (2) stripping - the potential is scanned anodically and Pb oxidises back, giving a sharp peak whose area/height is proportional to concentration. Preconcentration gives detection limits down to ~10^-10 M. Quantitation is by standard addition.
Required Apparatus & Chemicals
Laboratory Apparatus
- (Dry-lab) Stripping voltammograms: sample and two standard additions
- Ruler
- Mercury-film electrode (concept)
- Deposition timer (concept)
- Magnetic stirrer (concept)
Chemical Reagents
- (Dry-lab) Water sample
- Pb2+ standard
- Acetate buffer
- Mercury film (concept)
Step-by-Step Practical Procedure
- Measure the Pb stripping peak height for the sample.
- Measure peak heights after each standard addition.
- Plot peak height versus added concentration.
- Extrapolate to zero signal to find the original concentration.
- Convert to ug/L (ppb) and compare with the 10 ppb WHO guideline.
- Explain the role of the deposition step.
Observations & Expected Results
Peak heights grow linearly with added Pb; extrapolation typically gives low-ppb levels for tap water.
Worked Stoichiometric Calculations
Problem: Sample peak = 12 nA; +10 ppb spike gives 32 nA. Find the sample concentration.
Working: Slope = (32-12)/10 = 2.0 nA/ppb. C = 12/2.0 = 6.0 ppb - below the 10 ppb WHO guideline.
Safety Precautions & Hazard Control
- Dry-lab only: real ASV uses mercury films handled by trained staff.
- In practice, control deposition time precisely; it sets sensitivity.
- Deaerate the cell to remove O2 interference.
Waste Disposal Protocol
No wet waste in the dry-lab version; real mercury waste to dedicated collection.
Viva Voce Oral Defense Questions & Answers
Q: Why is ASV so sensitive?
A: The deposition step preconcentrates the analyte from a large solution volume into a tiny electrode volume, amplifying the stripping signal by 100-1000x.
Q: Why standard addition?
A: It corrects for matrix effects on deposition efficiency in real water samples.
Q: What sets the deposition potential?
A: It must be negative enough to reduce Pb2+ (on the diffusion plateau) but not so negative that other ions or H+ interfere.