Amperometric Titration: Lead with Dichromate (Dry-Lab)
To work through an amperometric titration of Pb2+ with Cr2O7(2-) as a dry-lab exercise and locate the endpoint from the current-volume plot.
Theoretical Foundations
In amperometry the current at a fixed applied potential is measured during titration. Holding the rotating Pt electrode at a potential where only Pb2+ reduces: current falls linearly as Pb2+ precipitates (Pb2CrO4), reaching zero at Ve; excess dichromate is not reducible at that potential, so current stays near zero. The intersection of the two lines gives Ve. (Dry-lab: mercury-free treatment.)
Required Apparatus & Chemicals
Laboratory Apparatus
- (Dry-lab) Titration data table: current vs volume
- Graph paper / spreadsheet
- Rotating platinum electrode (concept)
- DC power supply (concept)
- Nitrogen purge line (concept)
Chemical Reagents
- (Dry-lab) Pb2+ sample
- K2Cr2O7 titrant
- Supporting electrolyte
Step-by-Step Practical Procedure
- Tabulate the given current-volume data.
- Plot current versus titrant volume.
- Fit straight lines to the falling and flat regions.
- Read Ve at their intersection.
- Calculate [Pb2+] from the 2:1 Pb:Cr2O7 stoichiometry (2 Pb2+ + Cr2O7(2-) + H2O -> 2 PbCrO4 + 2 H+).
- Discuss why the current is zero after Ve.
Observations & Expected Results
Current drops linearly to near zero at Ve, then remains flat; Ve is read precisely at the intersection.
Worked Stoichiometric Calculations
Problem: 25.0 mL Pb2+ sample: Ve = 18.4 mL of 0.0200 M K2Cr2O7. Find [Pb2+].
Working: Moles Cr2O7 = 0.0184 x 0.0200 = 3.68 x 10^-4 mol; moles Pb2+ = 2 x 3.68 x 10^-4 = 7.36 x 10^-4 mol. [Pb2+] = 7.36 x 10^-4/0.0250 = 0.0294 M.
Safety Precautions & Hazard Control
- Dry-lab exercise; real lead and chromium(VI) compounds are toxic and need supervised handling.
- In practice, fix the applied potential on the diffusion plateau of Pb2+ only.
- Deaerate real amperometric cells.
Waste Disposal Protocol
No wet waste in the dry-lab version; real Pb/Cr waste to heavy-metal collection.
Viva Voce Oral Defense Questions & Answers
Q: Why is the current zero after the endpoint?
A: All Pb2+ is precipitated and dichromate is not reduced at the chosen potential, so no electroactive species remains.
Q: Why is amperometry good for precipitation titrations?
A: The endpoint comes from the current, so precipitates that obscure visual indicators cause no problem.