CHEMVEN

Polarography Concept: Half-Wave Potentials of Metal Ions (Dry-Lab)

To interpret polarographic waves of Cd2+, Pb2+ and Zn2+ as a dry-lab exercise: read E1/2 and limiting currents and apply the Ilkovic equation.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To interpret polarographic waves of Cd2+, Pb2+ and Zn2+ as a dry-lab exercise: read E1/2 and limiting currents and apply the Ilkovic equation.

Theoretical Foundations

In DC polarography the dropping mercury electrode gives a sigmoidal wave; the half-wave potential E1/2 identifies the ion (E1/2 = E0 + (0.05916/n) log(D_red/D_ox)^1/2, independent of concentration), while the diffusion-limited current i_d is proportional to concentration via the Ilkovic equation: i_d = 607 n D^1/2 m^2/3 t^1/6 C. Dissolved O2 is removed by N2 purging since it gives two reduction waves.

Required Apparatus & Chemicals

Laboratory Apparatus

  • (Dry-lab) Polarograms of Cd2+, Pb2+, Zn2+ at several concentrations
  • Ruler
  • Calculator
  • Nitrogen cylinder with regulator (deaeration, concept)
  • Mercury waste container (concept)

Chemical Reagents

  • (Dry-lab) Cd2+, Pb2+, Zn2+ solutions
  • KCl supporting electrolyte
  • Nitrogen (deaeration)

Step-by-Step Practical Procedure

  1. Measure E1/2 of each wave from the provided polarograms.
  2. Match each E1/2 to the ion using standard potential data.
  3. Measure limiting currents i_d at each concentration.
  4. Plot i_d vs C for Cd2+ and verify linearity through the origin.
  5. Use the Ilkovic equation with given capillary constants to compute D for Cd2+.
  6. Explain why N2 purging is essential.

Observations & Expected Results

E1/2 values (vs SCE): Pb2+ ~-0.40 V, Cd2+ ~-0.60 V, Zn2+ ~-1.00 V; i_d is linear in C for each ion.

Worked Stoichiometric Calculations

Problem: Cd2+ (n = 2): i_d = 4.84 uA at C = 1.00 mM; capillary gives m^2/3 t^1/6 = 1.50 mg^2/3 s^1/6. Calculate D (cm2/s).

Working: Ilkovic: i_d = 607 n D^1/2 m^2/3 t^1/6 C -> 4.84 = 607 x 2 x D^1/2 x 1.50 x 1.00. D^1/2 = 4.84/1821 = 2.658 x 10^-3; D = 7.06 x 10^-6 cm2/s.

Safety Precautions & Hazard Control

  • Dry-lab only: real polarography uses toxic mercury and must be done by trained staff in a fume hood.
  • Always deaerate real samples; O2 waves mask analyte waves.
  • Never handle mercury with bare hands.

Waste Disposal Protocol

No wet waste in the dry-lab version; real mercury waste goes to dedicated mercury collection.

Viva Voce Oral Defense Questions & Answers

Q: Why is E1/2 independent of concentration?

A: It depends only on E0 and the diffusion-coefficient ratio (Heyrovsky-Ilkovic equation), not on bulk concentration.

Q: Why purge with nitrogen?

A: Dissolved O2 reduces in two waves (~-0.1 and -0.9 V vs SCE) that overlap and obscure analyte waves.

Q: What does the Ilkovic equation tell you?

A: The diffusion current is proportional to n, sqrt(D), capillary characteristics and concentration - the basis of quantitative polarography.