CHEMVEN

Potentiometric Karl Fischer Concept: Water in a Solvent (Dry-Lab)

To understand the Karl Fischer determination of water by working through a potentiometric (bivoltametric) endpoint calculation as a dry-lab exercise.

Level: UniversitySafety: SafeTime: 60 min
Aim of Experiment: To understand the Karl Fischer determination of water by working through a potentiometric (bivoltametric) endpoint calculation as a dry-lab exercise.

Theoretical Foundations

Karl Fischer reaction: I2 + SO2 + H2O + 3 base -> 2 base-HI + base-HSO4 (1:1 I2:H2O). In the bivoltametric endpoint, a small constant current flows between two Pt electrodes; before Ve, I2/I- depolarises them (low voltage); after Ve, excess I2 keeps them polarised and the voltage jumps. Titre (mg H2O per mL reagent) converts the burette reading to water mass: % H2O = V x T / sample mass.

Required Apparatus & Chemicals

Laboratory Apparatus

  • (Dry-lab) Karl Fischer titrator schematic
  • Bivoltametric electrode pair diagram
  • Calculator
  • Drying oven diagram (glassware drying, concept)
  • Analytical balance (concept)

Chemical Reagents

  • (Dry-lab) Methanol solvent
  • KF reagent (titre given)
  • Water standard (sodium tartrate dihydrate, 15.66% H2O)

Step-by-Step Practical Procedure

  1. State the Karl Fischer reaction and its 1:1 stoichiometry.
  2. Explain the bivoltametric endpoint detection.
  3. Standardise the reagent conceptually with sodium tartrate dihydrate.
  4. Given a titre T = 5.00 mg H2O/mL, compute % water for a sample titration.
  5. List interferences (aldehydes, ketones, strong oxidants/reductants).
  6. Compare volumetric vs coulometric KF and state when each is used.

Observations & Expected Results

Endpoint: sudden voltage rise between the Pt electrodes on excess iodine; sodium tartrate dihydrate checks the titre at 15.66% water.

Worked Stoichiometric Calculations

Problem: 2.504 g of oil needed 3.20 mL of KF reagent (T = 5.00 mg/mL). Calculate % water.

Working: Water = 3.20 x 5.00 = 16.0 mg = 0.0160 g. % = 0.0160/2.504 x 100 = 0.639%.

Safety Precautions & Hazard Control

  • Dry-lab only: real KF reagents contain toxic SO2, iodine and methanol - fume hood mandatory in practice.
  • Keep samples and glassware dry; atmospheric moisture is the main error.
  • Interfering carbonyls need special reagents.

Waste Disposal Protocol

No wet waste in the dry-lab version; real KF waste goes to halogenated organic waste.

Viva Voce Oral Defense Questions & Answers

Q: Write the Karl Fischer reaction.

A: I2 + SO2 + H2O + 3 C5H5N -> 2 C5H5NHI + C5H5NSO3 (base = pyridine/imidazole); 1 mole I2 per mole H2O.

Q: How is the endpoint detected?

A: Bivoltametrically: a constant micro-current between two Pt electrodes; excess I2 after Ve polarises the electrodes and the voltage jumps.

Q: Volumetric vs coulometric KF?

A: Volumetric uses a burette of iodine reagent (mg-level water); coulometric generates I2 electrolytically and suits trace (ug-level) water.

← Back to Experiments Directory