Potentiometric Titration of Iodide with Permanganate (Redox, Acidic)
To determine iodide by potentiometric titration with KMnO4 in acidic medium using a Pt electrode.
Theoretical Foundations
2 MnO4- + 10 I- + 16 H+ -> 2 Mn2+ + 5 I2 + 8 H2O. The Pt electrode follows the I2/I- couple before Ve and the MnO4-/Mn2+ couple after it. The potential jump at Ve is large because the two formal potentials differ widely (0.54 V vs 1.51 V). Starch is unnecessary since detection is potentiometric.
Required Apparatus & Chemicals
Laboratory Apparatus
- mV meter
- Pt electrode
- Reference electrode
- Burette
- Beaker
- Stirrer
Chemical Reagents
- KI sample (~0.05 M)
- KMnO4 standard (0.02 M)
- 2 M H2SO4
Step-by-Step Practical Procedure
- Pipette 25.0 mL iodide sample, add 2 M H2SO4 to ~75 mL.
- Record the initial potential with Pt vs reference.
- Titrate with 0.02 M KMnO4, recording E.
- Take small steps through the jump.
- Plot E and dE/dV; read Ve.
- Calculate iodide molarity using the 1:5 MnO4-:I- stoichiometry.
Observations & Expected Results
The solution darkens (I2) then the brown colour fades near Ve; E jumps ~500 mV at the endpoint.
Worked Stoichiometric Calculations
Problem: 25.0 mL sample needed 20.4 mL of 0.0200 M KMnO4. Find [I-].
Working: Moles MnO4- = 0.0204 x 0.0200 = 4.08 x 10^-4 mol. Moles I- = 5 x 4.08 x 10^-4 = 2.04 x 10^-3 mol. [I-] = 2.04 x 10^-3/0.0250 = 0.0816 M.
Safety Precautions & Hazard Control
- Maintain strong acidity or MnO2 precipitates instead of Mn2+ forming.
- Standardise KMnO4 before use.
- Iodine vapours: keep the beaker covered loosely.
Waste Disposal Protocol
Iodine-containing acidic waste to halogenated aqueous waste; reduce MnO4- residues first.
Viva Voce Oral Defense Questions & Answers
Q: Write the balanced redox equation.
A: 2 MnO4- + 10 I- + 16 H+ -> 2 Mn2+ + 5 I2 + 8 H2O.
Q: Why must the medium be strongly acidic?
A: The MnO4-/Mn2+ couple needs 8 H+ per MnO4-; in weak acid, brown MnO2 forms instead and stoichiometry is lost.