Gran Plot Method: Linearising a Potentiometric Titration

To locate the equivalence point of a strong acid-strong base titration with a Gran plot, using data before the endpoint only.

Level: UniversitySafety: SafeTime: 90 min
Aim of Experiment: To locate the equivalence point of a strong acid-strong base titration with a Gran plot, using data before the endpoint only.

Theoretical Foundations

A Gran plot linearises the titration curve: for acid titrated with base, plot V_b x 10^-pH versus V_b (before Ve). Since [H+] ~ (C_a V_a - C_b V_b)/(V_a + V_b), the function is linear in V_b and extrapolates to zero at Ve. The method uses only pre-endpoint data, avoiding the unstable jump region, and is ideal for very dilute titrations.

Required Apparatus & Chemicals

Laboratory Apparatus

  • pH meter
  • Burette
  • Beaker
  • Stirrer
  • Pipette

Chemical Reagents

  • HCl (~0.01 M, dilute)
  • NaOH standard (0.01 M)
  • Calibration buffers

Step-by-Step Practical Procedure

  1. Calibrate the pH meter carefully.
  2. Pipette 50.0 mL of ~0.01 M HCl and dilute to ~100 mL.
  3. Titrate with 0.01 M NaOH, recording pH after each 1 mL addition, stopping before the jump.
  4. Compute the Gran function F = (V0 + Vb) x 10^-pH for each point.
  5. Plot F versus Vb; fit a straight line.
  6. Extrapolate to F = 0 to read Ve.
  7. Calculate [HCl] and compare with the derivative method.

Observations & Expected Results

The Gran plot is linear (r > 0.999); extrapolated Ve matches the derivative endpoint within 0.5%.

Worked Stoichiometric Calculations

Problem: Gran line: F = 4.20 x 10^-3 - 1.68 x 10^-4 Vb (Vb in mL). Find Ve.

Working: Set F = 0: Vb = 4.20 x 10^-3/1.68 x 10^-4 = 25.0 mL.

Safety Precautions & Hazard Control

  • Accurate pH values are essential; the antilog amplifies pH errors.
  • Use only pre-endpoint points well before the jump.
  • Calibrate meticulously at low concentrations.

Waste Disposal Protocol

Dilute neutralised solutions to drain.

Viva Voce Oral Defense Questions & Answers

Q: What is the advantage of the Gran method?

A: It uses only data far from the endpoint, where readings are stable, and linear extrapolation locates Ve objectively - especially useful for dilute titrations with small jumps.

Q: Why does the Gran function go to zero at Ve?

A: At the equivalence point [H+] (hence 10^-pH) approaches zero, so the product with volume extrapolates to zero.