Potentiometric Determination of Chloride in Table Salt (Purity Assay)
To assay the NaCl content of a commercial table-salt sample by potentiometric titration with AgNO3.
Theoretical Foundations
Ag+ + Cl- -> AgCl(s), 1:1 stoichiometry. A silver indicator electrode tracks [Ag+] via the Nernst equation; the dE/dV peak marks the endpoint. Mass of NaCl = V_AgNO3 x M_AgNO3 x 58.44 g/mol. Anti-caking agents do not interfere potentiometrically.
Required Apparatus & Chemicals
Laboratory Apparatus
- mV meter
- Silver electrode
- Reference electrode
- Burette
- Analytical balance
- Beaker
Chemical Reagents
- Table salt sample
- AgNO3 standard (0.1 M)
- Distilled water
Step-by-Step Practical Procedure
- Weigh ~0.25 g of table salt accurately and dissolve in ~75 mL water.
- Insert the silver and reference electrodes.
- Titrate with 0.1 M AgNO3, recording E in small steps near the jump.
- Plot E and dE/dV vs volume; read Ve.
- Calculate mass of NaCl and % purity.
- Repeat with a duplicate sample.
Observations & Expected Results
A sharp ~200 mV jump marks the endpoint; typical table salt assays 97-99.5% NaCl.
Worked Stoichiometric Calculations
Problem: 0.2512 g sample needed 42.6 mL of 0.100 M AgNO3. Calculate % NaCl.
Working: Moles Ag+ = 0.0426 x 0.100 = 4.26 x 10^-3 mol = moles NaCl. Mass = 4.26 x 10^-3 x 58.44 = 0.249 g. % = 0.249/0.2512 x 100 = 99.1%.
Safety Precautions & Hazard Control
- Weigh accurately to 0.1 mg.
- Protect AgNO3 from light.
- Stir to prevent AgCl coating the electrode.
Waste Disposal Protocol
Silver waste to heavy-metal collection.
Viva Voce Oral Defense Questions & Answers
Q: Why is the result expressed via the 1:1 stoichiometry?
A: Each Ag+ precipitates exactly one Cl-, so moles of AgNO3 at Ve equal moles of chloride, converted to NaCl mass with M = 58.44.
Q: Would iodide impurity interfere?
A: Yes: AgI precipitates before AgCl (lower Ksp), so iodide would be co-titrated and inflate the apparent chloride.