Conductometric Precipitation Titration: Chloride vs Silver Nitrate
To determine chloride by conductometric titration with AgNO3, following the fall and rise of conductance around the precipitation endpoint.
Theoretical Foundations
Ag+ + Cl- -> AgCl(s). Before the endpoint, added Ag+ removes Cl- and the highly mobile ions are replaced by slower NO3-, so conductance falls. After the endpoint, excess Ag+ and NO3- accumulate and conductance rises. The intersection of the two lines marks complete precipitation. The method needs no indicator and works in coloured samples.
Required Apparatus & Chemicals
Laboratory Apparatus
- Conductivity meter with cell
- Burette (amber or foil-wrapped for AgNO3)
- Beaker
- Stirrer
- Pipette
Chemical Reagents
- NaCl (~0.05 M)
- AgNO3 standard (0.1 M)
- Distilled water
Step-by-Step Practical Procedure
- Pipette 25.0 mL NaCl solution, dilute to ~100 mL.
- Record the initial conductance.
- Add 0.1 M AgNO3 in 1 mL portions with vigorous stirring, recording conductance.
- Continue 5 mL past the expected endpoint.
- Plot conductance vs volume and find the intersection of the two lines.
- Calculate the chloride molarity.
Observations & Expected Results
Conductance decreases linearly as AgCl precipitates, reaches a minimum at the endpoint, then increases as excess AgNO3 accumulates.
Worked Stoichiometric Calculations
Problem: Ve = 12.1 mL of 0.100 M AgNO3 for 25.0 mL sample. Calculate [Cl-].
Working: Moles Ag+ = 0.0121 x 0.100 = 1.21 x 10^-3 mol = moles Cl-. [Cl-] = 1.21 x 10^-3/0.0250 = 0.0484 M.
Safety Precautions & Hazard Control
- Protect AgNO3 from light; it decomposes and stains.
- Stir vigorously so the precipitate does not coat the electrodes.
- Keep temperature constant.
Waste Disposal Protocol
Silver chloride residues to heavy-metal waste; do not pour silver down the drain.
Viva Voce Oral Defense Questions & Answers
Q: Write the precipitation reaction and explain the conductance fall.
A: Ag+ + Cl- -> AgCl(s); ions leave the solution as solid, and remaining NO3- is less mobile than the Cl- removed, so conductance drops.
Q: Why is this better than the Mohr method for coloured samples?
A: No visual indicator is needed; the endpoint comes from the electrical measurement.