CHEMVEN

Paper Electrophoresis Concept: Separating Amino Acids by Charge

To separate amino acids by paper electrophoresis at a chosen pH and relate migration direction to their isoelectric points.

Level: UniversitySafety: Needs supervisionTime: 90 min
Aim of Experiment: To separate amino acids by paper electrophoresis at a chosen pH and relate migration direction to their isoelectric points.

Theoretical Foundations

In electrophoresis charged species migrate in an electric field; mobility depends on charge-to-size ratio. At a buffer pH below its pI an amino acid is cationic and moves toward the cathode; above its pI it is anionic and moves toward the anode; at pH = pI it stays at the origin. Aspartic acid (pI ~3.0), histidine (pI ~7.6) and lysine (pI ~9.7) therefore fan out at pH 6.

Required Apparatus & Chemicals

Laboratory Apparatus

  • Paper electrophoresis tank with electrodes
  • DC power supply
  • Filter paper strips
  • Ninhydrin spray
  • pH meter

Chemical Reagents

  • Aspartic acid, histidine, lysine solutions
  • Phosphate buffer pH 6.0
  • Ninhydrin in ethanol

Step-by-Step Practical Procedure

  1. Soak paper strips in pH 6 buffer and mount them in the tank.
  2. Spot the three amino acids at the marked origin line.
  3. Apply about 200 V for 45-60 min, keeping the apparatus cool.
  4. Switch off, remove the strips and dry them.
  5. Spray with ninhydrin and heat to develop the spots.
  6. Measure migration distances and directions relative to the origin.

Observations & Expected Results

At pH 6 lysine (pI 9.7, cationic) moves toward the cathode, aspartic acid (pI 3.0, anionic) toward the anode, and histidine (pI 7.6) moves only slightly toward the cathode.

Worked Stoichiometric Calculations

Problem: Lysine migrated 4.0 cm toward the cathode in 45 min. What is its electrophoretic mobility?

Working: Velocity = 4.0 cm / 2700 s = 1.48 x 10^-3 cm/s. Mobility = velocity/field; with a 20 cm strip at 200 V, field = 10 V/cm, so mobility = 1.48 x 10^-4 cm^2/Vs.

Safety Precautions & Hazard Control

  • High voltage: never touch the buffer or electrodes while the power is on.
  • Keep the tank covered and cool to limit Joule heating and evaporation.
  • Spray ninhydrin in a fume hood.

Waste Disposal Protocol

Buffer to drain; ninhydrin-stained paper to solid chemical waste.

Viva Voce Oral Defense Questions & Answers

Q: Define the isoelectric point.

A: The pH at which the amino acid carries no net charge and does not migrate in an electric field.

Q: Predict the migration of aspartic acid at pH 6 and explain.

A: Toward the anode, because pH 6 is above its pI (~3.0) so it is negatively charged.

Q: Why is the apparatus cooled?

A: Joule heating causes evaporation, buffer concentration changes and band broadening, which spoil the separation.

← Back to Experiments Directory