Gel Filtration (Size Exclusion): Desalting a Protein Sample
To desalt a coloured protein solution on a Sephadex G-25 column, separating the macromolecule from low-molecular-weight salt.
Theoretical Foundations
In size-exclusion chromatography porous beads fractionate by molecular size: large molecules are excluded from the pores and elute in the void volume V0, while small molecules penetrate the pores and elute later, near the total volume Vt. A coloured high-MW marker (e.g. blue dextran for V0) and a small dye or salt (for Vt) bracket the separation; the protein elutes with the void peak, free of salt.
Required Apparatus & Chemicals
Laboratory Apparatus
- Sephadex G-25 column
- Fraction tubes
- Conductivity meter or AgNO3 test
- Spectrophotometer (280 nm)
- Measuring cylinder
Chemical Reagents
- Protein sample in 1 M NaCl (e.g. coloured albumin or haemoglobin)
- Blue dextran solution (void marker)
- Distilled water / buffer eluent
- AgNO3 test solution
Step-by-Step Practical Procedure
- Pack and equilibrate the Sephadex column, determining V0 with blue dextran.
- Load the protein-in-NaCl sample carefully onto the bed.
- Elute with water or dilute buffer, collecting 2 mL fractions.
- Test each fraction: protein by A280 or colour, salt by conductivity or AgNO3.
- Plot protein signal and salt signal versus fraction number.
- Pool the early protein peak (salt-free) and the late salt peak separately.
Observations & Expected Results
The coloured protein peak emerges first at the void volume, well before the conductivity/AgCl salt peak; pooled protein fractions give a negative chloride test.
Worked Stoichiometric Calculations
Problem: Blue dextran eluted at 12 mL (V0) and NaCl peaked at 34 mL. The column bed volume Vt is 40 mL. What is the distribution coefficient Kav of NaCl?
Working: Kav = (Ve - V0)/(Vt - V0) = (34 - 12)/(40 - 12) = 22/28 = 0.79, typical for a fully included small molecule.
Safety Precautions & Hazard Control
- Never let a gel-filtration bed run dry; the gel cracks and resolution is lost.
- Load the sample as a narrow band for sharp peaks.
- Use particle-free eluents to avoid blocking the bed.
Waste Disposal Protocol
Protein and salt fractions are non-hazardous and can go down the drain; used gel to solid waste.
Viva Voce Oral Defense Questions & Answers
Q: Why does the protein elute before the salt?
A: The protein is too large to enter the gel pores so it travels only in the void volume, while NaCl diffuses into the pores and takes a longer path.
Q: What are V0 and Vt?
A: V0 is the void (excluded) volume eluted by totally excluded molecules; Vt is the total bed volume marking elution of fully included small molecules.
Q: Why is this called desalting?
A: The macromolecule is recovered in the early fractions essentially free of the low-molecular-weight salt, which elutes much later.