Beer–Lambert Law
Absorbance is proportional to concentration and path length: A = εbc.
Overview & Foundations
Start with the measurement itself. A purple potassium permanganate solution sits in a 1.00 cm cuvette, and the spectrophotometer reports an absorbance of 0.550 at the wavelength where its molar absorptivity ε is 1.10×10⁴ L mol⁻¹ cm⁻¹. One line of arithmetic turns that reading into a concentration: c = A ÷ (εl) = 0.550 ÷ (1.10×10⁴ × 1.00) = 5.00×10⁻⁵ mol/L. No titration, no weighing of a precipitate — just light, a logarithm and a constant. That calculation, the Beer–Lambert law in action, is how an enormous share of quantitative analysis is actually done, so it is worth understanding exactly what each symbol is hiding.
The law grew in three steps. Pierre Bouguer and then Johann Lambert showed that each successive layer of a material absorbs the same fraction of the light passing through it, so intensity falls exponentially with path length, not linearly. August Beer added, in 1852, that the same exponential fall comes from concentration: double the absorbers, double the effect per layer. Because the fall is exponential, analysts work with its logarithm. Transmittance T is the fraction of light that gets through, I/I₀; absorbance is defined as A = −log₁₀(T). That one definition is what straightens an exponential decay into the simple proportion A = εcl.
Read the law symbol by symbol. A is absorbance, a pure number with no units. l (often written b) is the path length — the distance light travels through the solution, almost always the 1.00 cm width of a standard cuvette. c is concentration in mol/L. And ε, the molar absorptivity, is the molecule's own property at a chosen wavelength: how strongly one mole per litre absorbs across one centimetre. A large ε, like the 1.10×10⁴ of the example, means an intensely absorbing species that can be measured at very low concentration; a small ε means a feeble absorber needing concentrated solutions. Note what the law does not contain: the colour itself. Measurements are made at one selected wavelength, normally the absorption maximum λmax — for permanganate the strong band at 525–545 nm — because there the sensitivity is greatest and small wavelength errors matter least.
In practice the law is used through a calibration line. Prepare standard solutions of known concentration, measure each absorbance at λmax in the same cuvette, and plot A against c: the law promises a straight line through the origin with gradient εl. Then read the unknown's absorbance off against the line. Using standards measured on the same instrument, on the same day, quietly cancels most instrumental imperfections — which is why calibration, rather than a textbook ε value alone, is the working analyst's method.
The law states its own limits if you read the table of A against %T. At A = 1.0 only 10% of the light reaches the detector; at A = 2.0 only 1% does, so tiny amounts of stray light leaking around the sample, or detector noise, become a large fraction of the signal and the calibration line bends towards the concentration axis. Chemical deviations bend it too: at high concentration molecules can dimerise, or an indicator can shift its acid–base equilibrium, changing the absorbing species itself. Scattering from a cloudy sample fakes absorbance that no molecule produced. The honest working range on most instruments is therefore up to about A ≈ 1 to 1.5; stronger samples are diluted into range rather than trusted at face value. Polychromatic light — a poorly chosen, wide wavelength band — flattens peaks and breaks the linearity for the same reason: ε is only constant at one wavelength.
Good practice is mostly respect for those limits. Blank the instrument with the pure solvent in the same cuvette so solvent and cell absorption cancel; keep cuvettes clean, matched and filled to the same level, with the clear faces towards the beam; choose λmax; dilute anything reading above the linear range and multiply back by the dilution factor afterwards.
So the worked example at the top generalises into a complete method: light dims exponentially, the logarithm makes it linear, ε measures the molecule's appetite for photons at the chosen wavelength, and calibration turns readings into concentrations. Respect the linear range and the Beer–Lambert law is one of the most reliable bargains in analytical chemistry — a concentration from nothing but a beam of light.
Core Chemical Concepts
- A = εcl: absorbance is directly proportional to concentration and path length
- A = −log₁₀(T): the logarithm turns exponential dimming into a straight line
- ε, molar absorptivity, is the molecule's property at one chosen wavelength (L mol⁻¹ cm⁻¹)
- Measure at λmax — for KMnO₄ the 525–545 nm band — for greatest sensitivity
- Calibration line: plot A against c for standards; gradient = εl, line through the origin
- Linear only at moderate absorbance: above about A ≈ 1.5, stray light and chemistry bend the line
Governing Formulas & Laws
Beer–Lambert law: ε molar absorptivity, c concentration in mol/L, l path length in cmAbsorbance from transmittance: A = 1 means T = 0.10, only 10% of light transmittedThe working rearrangement: concentration from a measured absorbanceConverting absorbance back to percentage transmittanceWorked Numerical Examples
Industrial & Laboratory Applications
- Water quality testing: trace metal and nitrate concentrations measured by colorimetry in minutes
- Drug and clinical assays: concentrations of medicines and blood components from a single absorbance reading
- DNA purity checks: the A₂₆₀/A₂₈₀ absorbance ratio assesses nucleic acid samples in molecular biology
- Kinetics: following absorbance against time tracks concentration and yields reaction rates without sampling
Common Student Misconceptions
Everyday Life & Real-World Relevance
- A = εcl; rearrange to c = A ÷ (εl) for concentration questions
- Standard cuvette path length l = 1.00 cm; A has no units
- A = 1 → 10% T; A = 2 → 1% T; each unit of A cuts transmitted light tenfold
- Measure at λmax, blank with solvent, dilute samples that read too high
- Calibration line of A against c passes through the origin with gradient εl